Remote Control Commands
R&S
®
SMC100A
258
Operating Manual 1411.4060.32 ─ 04
Parameters:
<Manual>
float
Range:
STARt to STOP
Increment: 0.01Hz
*RST:
100MHz
Default unit: Hz
Example:
SWE:MODE MAN
sets the
Step
sweep mode.
Example:
FREQ:MAN 500MHz
sets an RF frequency of 500 MHz for the next step in the
Step
sweep mode.
FREQ:MODE SWE
sets the Frequency Sweep mode. An RF frequency of 500 MHz is
output.
FREQ:MAN 550MHz
triggers the next sweep step with an RF frequency of 550 MHz.
Usage:
SCPI confirmed
[:SOURce<hw>]:FREQuency:MODE
<Mode>
The command sets the instrument operating mode and therefore also the commands
used to set the output frequency.
Parameters:
<Mode>
CW | FIXed | SWEep
CW|FIXed
The instrument operates in fixed-frequency mode;
CW
and
FIXed
are synonyms. The output frequency is set
with
:SOURce:FREQuency:CW|FIXed
.
SWEep
The instrument operates in SWEep mode. The frequency is set
using the commands
SOURce:FREQuency:STARt; STOP;
CENTer; SPAN; MANual
.
*RST:
CW
Example:
FREQ:MODE SWE
sets the SWEep mode. The settings under
SOURce:FREQuency:STARt
;
STOP
;
CENTer
;
SPAN
;
MANual
become effective.
Usage:
SCPI confirmed
[:SOURce<hw>]:FREQuency:MULTiplier
<Multiplier>
The command sets the value for the multiplication factor of a subsequent downstream
instrument.
SOURce Subsystem