Remote Control Commands
R&S
®
SMBV100B
925
User Manual 1178.4460.02 ─ 03
:UNIT:ANGLe
<Angle>
Sets the default unit for phase modulation angle. The command affects no other
parameters, such as RF phase, or the manual control or display.
Parameters:
<Angle>
DEGree | DEGRee | RADian
*RST:
RADian
:UNIT:POWer
<Power>
Sets the default unit for all power parameters. This setting affects the GUI, as well as
all remote control commands that determine power values.
Parameters:
<Power>
V | DBUV | DBM
*RST:
DBM
:UNIT:VELocity
<Velocity>
Sets the default unit for the velocity of the wave.
Parameters:
<Velocity>
MPS | KMH | MPH | NMPH
*RST:
MPS
UNIT Subsystem