Remote Control Commands
R&S
®
SMBV100A
760
Operating Manual 1176.8016.02 ─ 17
Parameters:
<Mode>
CW | FIXed | SWEep
CW|FIXed
Sets the CW frequency mode.
CW
and
FIXed
are synonyms.
The instrument operates at a fixed frequency. To set the LF out-
put frequency use the command
SWEep
Sets the sweep mode.
The instrument processes the frequency settings in defined
sweep steps. To determine the corresponding frequency values,
use the commands
[:SOURce<hw>]:LFOutput:FREQuency:
[:SOURce<hw>]:LFOutput:FREQuency:STOP
or
SOURce<hw>]:LFOutput:FREQuency:MANual
*RST:
CW
Example:
LFO:FREQ:MODE SWE
sets the sweep mode.
Manual operation:
See
[:SOURce<hw>]:LFOutput:FREQuency:STARt
<Start>
Sets the start frequency for the LF sweep.
Parameters:
<Start>
float
Range:
full frequency range
Increment: see the data sheet: Resolution of frequency setting
*RST:
1 KHz
Example:
RST*
activates all presettings.
LFO:SWE:MODE AUTO
TRIG0:SOUR SING
LFO:FREQ:STAR 1 kHz
LFO:FREQ:STOP 10 kHz
LFO:FREQ:MODE SWE
LFO:SWE:EXEC
the instrument generates a single sweep cycle from 1 kHz to 10
kHz automatically after a manual trigger event occurs
(
:LFOutput:SWEep:EXECute
or
*TRG
). The step width is 1
kHz linear, with 15 ms dell time until the signal switches to the
subsequent step.
Manual operation:
See
[:SOURce<hw>]:LFOutput:FREQuency:STOP
<Stop>
Sets the stop frequency for the LF sweep.
SOURce Subsystem