Remote Control Commands
R&S
®
SMBV100A
516
Operating Manual 1176.8016.02 ─ 17
Example:
{TYPE:SMU-WV, 837236424}
{COPYRIGHT:2003 Rohde&Schwarz SMU}
{DATE:2012-07-11;14:38:01}
{SAMPLES:80000}
{CLOCK:8666666.666666666}
{VECTOR MAX:1.000000038569158}
{LEVEL OFFS:3.333553817875577e-07,0}
{MARKER LIST 1:0:1;1:0;1249:0}
{MARKER LIST 2:0:1;1:0;1249:0}
{MARKER LIST 3:0:1;1:0;1249:0}
{MARKER LIST 4:0:1;1:0;1249:0}
{EMPTYTAG-15947:# ...}
{WAVEFORM-320017:#IQIQIQ...}
The example waveform file contains 436 (0x1b4) bytes before
the
#
sign in the
EMPTYTAG
; the hex address of the
#
sign is
0x1b5
. The
{WAVEFORM}
starts at
0x4000
. The
EMPTYTAG
con-
tains
15946
blanks and has a length of (15946+1) bytes.
Usage:
Setting only
{CONTROL LENGTH: ControlLength}
(optional / recommended for marker and control lists)
The tag specifies the length of
all
control or marker list in ASCII format.
The control length influences the way the marker and control lists are processed, in
particular the way traces are repeated; see
Figure 7-1: Example: Processing of MARKER TRACE if CONTROL LENGHT is specified
If the
CONTROL LENGTH
tag is not used, the marker and control list length are deter-
mined by the last position, that is the last {
Pos
:State} couple, defined in the particular
[TRACE] LIST
tag; see
.
Figure 7-2: Example: Processing of MARKER TRACE if CONTROL LENGHT is not used
Waveform and List Format