![Rohde & Schwarz SMBV100A Скачать руководство пользователя страница 262](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_operating-manual_1477735262.webp)
Instrument Function
R&S
®
SMBV100A
247
Operating Manual 1176.8016.02 ─ 17
"User"
The filter length is set manually.
The filter length is entered in the entry window to the right. As the fil-
ter length works as a multiplier for the time window, this results in a
constant measurement time.
Note:
The time window varies depending on the used sensor. For
most sensors it is fixed to 20 ms. For the R&S NRP-Z81 sensor it is
10 us. Therefore, the user filter length for the R&S NRP-Z81 has be
about 1000 times larger than the filter length for other sensors in
order to achieve the same filtering result.
The
button can be used to search for the optimum filter
length for the current measurement conditions. The found filter length
is indicated in the field to the right, see
"Fixed Noise"
The averaging factor is selected so that the sensors intrinsic noise (2
standard deviations) does not exceed the specified noise content.
The desired noise content is entered in the entry field to the right, see
To avoid very long settling times when the power is low, the averag-
ing factor can be limited with the
Remote command:
:SENSe<ch>[:POWer]:FILTer:TYPE
Filter Length ← Filter
Indicates the used filter length for filter type "Auto" or "User".
Remote command:
:SENSe<ch>[:POWer]:FILTer:LENGth:AUTO?
:SENSe<ch>[:POWer]:FILTer:LENGth[:USER]
Noise Content ← Filter
Sets the noise content for filter type "Fixed Noise".
Remote command:
:SENSe<ch>[:POWer]:FILTer:NSRatio
Timeout ← Filter
Sets a time limit for the averaging process.
Remote command:
:SENSe<ch>[:POWer]:FILTer:NSRatio:MTIMe
Auto Once ← Filter
Calculates the optimum filter length for the current measurement conditions and indi-
cates the value in the
.
Remote command:
:SENSe<ch>[:POWer]:FILTer:SONCe
Use Default Aperture Time
Enables you to specify a user-defined aperture time for the respective sensor.