Remote Control Commands
R&S
®
SMBV100A
487
Operating Manual 1407.6062.32 ─ 08
Example:
SENS:UNIT DBM
selects unit dBm for presentation of measurement result.
READ1?
queries the measurement result of the sensor connected to the
SENSOR interface.
Response: -45.6246576745440230
-45.6 dBm were measured at the given frequency.
or e.g. for R&S NRP-Z81
Response:
-55.62403263352178,-22.419472478812476
-55,6 dbm is the measured average level, -22. 4 dBm is the mea-
sured peak level at the given frequency
Usage:
Query only
SENSe<ch>[:POWer]:CORRection:SPDevice:STATe
<State>
The command activates the use of the s-parameters correction data of the selected power
sensor.
Note:
For power sensor with attenuator this command is automatically set to ON.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
OFF
Example:
SENS:POW:CORR:SPD:STAT ON
activates the use of the s-parameters correction data of power
sensor 1.
SENSe<ch>[:POWer]:DISPlay:PERManent:STATe
<State>
The command switches on and off the permanent indication of the power measurement
result in the upper right corner of the block diagram. For each sensor, the type of sensor,
the connector, the measurement source and - if set - the offset is indicated.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
OFF
Example:
SENS1:POW:DISP:PERM:STAT ON
the permanent viewer is switched on.
SENSe<ch>[:POWer]:DISPlay:PERManent:PRIority
<Priority>
The command selects which power measurement result (average or peak power) is indi-
cated when permanent display is active.
Parameters:
<Priority>
AVERage | PEAK
*RST:
AVERage
SENSe, READ and INITiate Subsystems
Содержание SMBV100A