Instrument Function
R&S
®
SMBV100A
383
Operating Manual 1407.6062.32 ─ 08
The delay can be defined by moving the setting mark.
SCPI command:
Fix marker delay to current range
Restricts the marker delay setting range to the dynamic range. In this range the delay
can be set without restarting the marker and signal.
SCPI command:
Clock Settings
The Clock Settings is used to set the clock source and a delay if required.
Sync. Mode
Selects the synchronization mode.
This parameter is used to enable generation of very precise synchronous signal of several
connected R&S SMBVs.
Note:
If several instruments are connected, the connecting cables from the master
instrument to the slave one and between each two consecutive slave instruments must
have the same length and type.
Avoid unnecessary cable length and branching points.
"None"
The instrument is working in stand-alone mode.
"Sync. Master" The instrument provides all connected instrument with its synchronisa-
tion (including the trigger signal) and reference clock signal.
"Sync. Slave"
The instrument receives the synchronisation and reference clock signal
from another instrument working in a master mode.
SCPI command:
Set Synchronization Settings
Performs automatically adjustment of the instrument's settings required for the synchro-
nization mode, selected with the parameter
.
SCPI command:
Clock Source
Selects the clock source.
"Internal"
The internal clock reference is used to generate the symbol clock.
Baseband Signal - Baseband Block
Содержание SMBV100A
Страница 31: ......
Страница 117: ...Manual Operation R S SMBV100A 100 Operating Manual 1407 6062 32 08 Legend of Front Panel Controls ...