Known-Issues
RF
If no K31 is installed, an error message appears for levels > 20 dBm
but does not disappear for levels < 20 dBm
735336
External Reference
In rare cases the error message 'RFB6.BB: Output unleveled: Gain
Control on Upper Limit' appears if the external reference frequency
was switched from external to internal.
421671
1.2 Version 4.70.006.33
Released: February 2020
New Options
None
New Functionality / Changed Behavior
Common
Context menu can be accessed by pressing the rotary knob for some
time.
362578
Fixed Issues
Common
Hardcopy
By default, screen-shots are not saved to USB memory
550938
Mouse operation disturbed after touching screen, remote control via
VNC does not work as expected.
489231
RF
In rare cases the error message RFB6.BB: Output unleveled: Gain
Control on Lower Limit appears, if pulse modulation is active.
585193
AM Modulation
The modualtion depth didn't change the PEP value.
510310
Stereocoder
Direct mode read back does not work
597784
Read back of PS does not work
599475
Pulse Generator
In rare cases the pulse train is wrong, if a zero pulse is added at the
end of the pulse train.
426320