Remote Control Commands
R&S
®
Scope Rider RTH
382
User Manual 1326.1578.02 ─ 12
Parameters:
<Range>
<numeric value> | 'AUTO' | MIN | MAX | DEF
<numeric value>
The instrument has fixed measurement ranges: 1 V|10 V|100 V|
1000
V.
You can enter any value between the minimum and maximum
value. The instrument uses the next suitable range.
Range:
1 to 1000
*RST:
1
Default unit: V
Usage:
Setting only
15.7.2.2
METer:SENSe:<function>:RANGe:UPPER Commands
METer:SENSe:<function>:RANGe:UPPER
commands set the measurement range
and turn off autoranging. You can use these commands in addition to
METer:CONFigure
commands if you want to change only the range.
METer:SENSe:CAPacitance:RANGe:UPPer
..................................................................... 382
METer:SENSe:CURRent:AC:RANGe:UPPer
.....................................................................382
METer:SENSe:CURRent:DC:RANGe:UPPer
.....................................................................382
METer:SENSe:FREQuency:VOLTage:RANGe:UPPer
........................................................ 383
METer:SENSe:RESistance:RANGe:UPPer
....................................................................... 383
METer:SENSe:VOLTage:AC:RANGe:UPPer
..................................................................... 383
METer:SENSe:VOLTage:DC:RANGe:UPPer
..................................................................... 383
METer:SENSe:CAPacitance:RANGe:UPPer
<Range>
Sets a fixed range for capacitance measurements and turns off auto ranging.
Parameters:
<Range>
<numeric value> | MIN | MAX | DEF
<numeric value>
The instrument has fixed measurement ranges: 10 nF|100 nF|
1
μF|10 μF|100 μF|1 mF|10 mF.
You can enter any value between the minimum and maximum
value. The instrument uses the next suitable range.
Range:
10 nF to 10 mF
*RST:
10 nF
METer:SENSe:CURRent:AC:RANGe:UPPer
<Range>
METer:SENSe:CURRent:DC:RANGe:UPPer
<Range>
Sets a fixed current range and turns off auto ranging.
Digital Multimeter (R&S
RTH1002)