Remote Control Commands
R&S
®
Scope Rider RTH
372
User Manual 1326.1578.02 ─ 12
HARMonic:EXPort:SAVE
Saves the results of the harmonic measurement to the file specified using
.
Example:
HARMonic:EXPort:NAME 'HarmonicResults'
HARMonic:EXPort:SAVE
HARMonic:EXPort:NAME?
//Result: /media/SD/Rohde-Schwarz/RTH/Export/HarmonicResults.csv
Usage:
Event
HARMonic:RESult<m>:CLIPping?
Indicates whether clipping has occurred, that is: the amplitudes exceed the currently
defined amplitude range.
Suffix:
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
Input channel
Return values:
<State>
ON | OFF
ON
Clipping has occurred, increase the amplitude range (see
OFF
No clipping has occurred, the current measurement settings are
suitable.
*RST:
OFF
Usage:
Query only
HARMonic:RESult<m>:FRQMissing?
Queries whether the fundamental frequency was detected in the input signal or not. If
the specified frequency
±
10
% is not detected in the signal, the measurement is invalid.
Suffix:
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
Return values:
<State>
ON | OFF
OFF
The fundamental frequency was not detected, the measurement
is invalid. Define the correct frequency of the input signal.
ON
The fundamental frequency was detected, the measurement is
valid.
*RST:
OFF
Spectrum Analysis