Remote Control Commands
R&S
®
Scope Rider RTH
380
User Manual 1326.1578.02 ─ 12
The
METer:CONFigure:<function>
commands are only for configuration. To acti-
vate a measurement, use
returns the active measurement and range.
To read the result, use
.
........................................................................................ 380
...............................................................................................380
..................................................................................... 380
...................................................................................... 380
..................................................................................... 380
..................................................................................... 380
....................................................................................... 381
........................................................................................381
...................................................................................... 381
......................................................................................381
METer:CONFigure:CONTinuity
METer:CONFigure:DIODe
METer:CONFigure:TEMPerature
Configures the specified measurement. The instrument sets a fixed range.
Usage:
Event
METer:CONFigure:CAPacitance
Sets the internal parameters and configures the range for capacitance measurements.
Parameters:
<Range>
<numeric value> | 'AUTO' | MIN | MAX | DEF
<numeric value>
The instrument has fixed measurement ranges: 10 nF|100 nF|
1
μF|10 μF|100 μF|1 mF|10 mF.
You can enter any value between the minimum and maximum
value. The instrument uses the next suitable range.
Range:
10 nF to 10 mF
*RST:
10 nF
Usage:
Setting only
METer:CONFigure:CURRent:AC
METer:CONFigure:CURRent:DC
Sets the internal parameters and configures the range for current measurements.
You need an external shunt resistor or I/U converter for current measurement.
Digital Multimeter (R&S
RTH1002)