Remote Control
R&S
®
RTM20xx
305
User Manual 1317.4726.02 ─ 01
TRIGger:A:EDGE:COUPling
<Coupling>
Sets the coupling for the trigger source.
Parameters:
<Coupling>
DC | AC | HF
DC
Direct Current coupling. The trigger signal remains unchanged.
AC
Alternating Current coupling. A 5 Hz high pass filter removes the
DC offset voltage from the trigger signal.
HF
High frequency coupling. A 15 kHz high-pass filter removes lower
frequencies from the trigger signal. Use this mode only with very
high frequency signals.
*RST:
DC
TRIGger:A:EDGE:FILTer:LPASs
<State>
Turns an additional 5 kHz low-pass filter in the trigger path on or off. This filter removes
higher frequencies and is available with AC and DC coupling.
Parameters:
<State>
ON | OFF
*RST:
OFF
TRIGger:A:EDGE:FILTer:NREJect
<State>
Turns an additional 100 MHz low-pass filter in the trigger path on or off. This filter removes
higher frequencies and is available with AC and DC coupling.
Parameters:
<State>
ON | OFF
*RST:
OFF
15.2.5.3
Width Trigger
............................................................................................305
..............................................................................................306
...............................................................................................306
...............................................................................................306
TRIGger:A:WIDTh:POLarity
<Polarity>
Sets the polarity of the pulse.
Command Reference