Protocol Analysis
R&S
®
RTM20xx
188
User Manual 1317.4726.02 ─ 01
Overload ← Frame <type>
An overload frame is sent by a node that needs a delay between data and/or remote
frames.
Data ← Frame <type>
Frame for data transmission.
The identifier format is also considered, see
Remote ← Frame <type>
Data: Frame for data transmission.
A remote frame initiates the transmission of data by another node. The frame format is
the same as of data frames but without the data field.
The identifier format is also considered, see
Data or remote ← Frame <type>
Triggers on remote frames and on data frames.
The identifier format is also considered, see
Error <type>
Identifies various errors in the frame. You can select one or more error types as trigger
condition.
The first key press enables the "Error" trigger type, the second opens the "Error" menu.
Remote command:
ERRCondition
)
Stuff bit ← Error <type>
The frame segments Start Of Frame, Arbitration Field, Control Field, Data Field and CRC
Sequence are coded by the bit stuffing method. The transmitter automatically inserts a
complementary bit into the bit stream when it detects five consecutive bits of identical
value in the bit stream to be transmitted. A stuff error occurs when the 6th consecutive
equal bit level in the mentioned fields is detected.
Remote command:
Form ← Error <type>
A form error occurs when a fixed-form bit field contains one or more illegal bits.
Remote command:
Acknowledge ← Error <type>
An acknowledgement error occurs when the transmitter does not receive an acknowl-
edgment - a dominant bit during the Ack Slot.
Remote command: