Remote Control Commands
R&S
®
RTM3000
460
User Manual 1335.9090.02 ─ 08
Parameters:
<DelayTime>
Range:
20e-9 to 6,871946854
Increment: Depends on the <DelayTime> value. The longer the
time, the higher the increment.
*RST:
20e-9
Default unit: s
TRIGger:B:EVENt:COUNt
EventCount
Sets a number of B-trigger events that fulfill the B-trigger condition but do not cause
the trigger. The oscilloscope triggers on the n-th event, the last of the specified number
of events.
Parameters:
EventCount
Range:
1 to 65535
Increment: 1
*RST:
1
16.5.4
Width Trigger
To set the trigger level (threshold), use:
●
●
CHANnel<m>:THReshold:FINDlevel
●
CHANnel<m>:THReshold:HYSTeresis
............................................................................................ 460
..............................................................................................461
...............................................................................................461
............................................................................................... 461
TRIGger:A:WIDTh:POLarity
<Polarity>
Sets the polarity of the pulse.
Parameters:
<Polarity>
POSitive | NEGative
POSitive
Positive going pulse, the width is defined from the rising to the
falling slopes.
NEGative
Negative going pulse, the width is defined from the falling to the
rising slopes.
*RST:
POSitive
Trigger