R&S RTM Digital Oscilloscope
Installation Information
Release Notes 1305.1004.00 - 11
4
B10 option)
RTM_Release_Notes_V4.600.pdf
This document
Extract these files from the archive to a memory stick.
2.1.2 Performing the Firmware Update on the Instrument
The firmware update process is performed by the following steps:
1. Insert the memory stick into the USB port of the oscilloscope and push the key
Setup
.
2. Press the soft key
More 1|3
.
3. Press the soft key
Update
.
4. Press the soft key
Firmware
.
5. Press the soft key
Execute
if the new firmware loaded and checked.
6. Wait until the device reboot automatically.
The oscilloscope will update the firmware and reboot itself. Please make sure, that
there is no interrupt at the main power during the update process. Otherwise this can
be make the instrument unusable.
2.1.3 Performing the Firmware Update on the Interface Board with USB
and LAN
The firmware update process for the interface board is performed by the following
steps (steps 1 to 3 could be skipped since interface board is already installed):
1. Turn off the instrument.
2. Install the interface board with the USB and LAN interface in the instrument (rear
side of the instrument).
3. Turn on the instrument.
4. Insert the memory stick into the USB port of the oscilloscope and push the key
Setup
.
5. Press the soft key
More 1|3
.
6. Press the soft key
Update
.
7. Press the soft key
Interface
.
8. Press the soft key
Execute
if the new firmware loaded and checked.
9. Wait until the update process is finished.