Waveform setup
R&S
®
RTB2000
69
User Manual 1333.1611.02 ─ 11
"Enve-
lope
+
PD"
Each acquisition is done in peak detect mode, and the most extreme
values of all consecutive acquisitions build the envelope. This method
is more precise than "Envelope".
Remote command:
No. of Averages
Defines the number of waveforms used to calculate the average waveform. The higher
the number, the better the noise is reduced.
To restart the average calculation, press the [Clear Screen] key.
Remote command:
Nx Single
Sets the number of waveforms that are acquired with a [Single] acquisition.
The setting is available if the history option is installed.
Remote command:
Roll
Enables the automatic roll mode. The instrument switches to roll mode if the
The roll mode displays the untriggered, continuous signal, and moves the captured
input data on the display from the left to the right. The instrument shows the waveform
immediately, without waiting for the complete acquisition of the waveform record. The
record length is set automatically ("Auto" mode). Some math functions are non-calcula-
ble if roll mode is active.
You can use the horizontal and vertical zoom in roll mode if the acquisition is stopped.
Remote command:
Start Roll Time
Sets the limit timebase for the roll mode. The instrument switches automatically to roll
mode if:
●
the
exceeds the value given here.
●
the roll mode is activated (
Remote command:
Acquisition setup