Remote control commands
R&S
®
RTB2000
461
User Manual 1333.1611.02 ─ 11
SYSTem:COMMunicate:INTerface:USB:CLASs
Selects the USB mode.
●
USB TMC (Test & Measurement Class)
●
USB VCP (Virtual Com Port)
●
USB MTP (Media Transfer Protocol)
Parameters:
<USBClass>
TMC | VCP | MTP
15.10.5
Trigger out
TRIGger:OUT:MODE
<OutputMode>
Defines which signals are generated at the [Aux
Out] connector.
Parameters:
<OutputMode>
OFF | TRIGger | REFerence | MASK | GENerator
OFF
No output
TRIGger
Outputs a pulse when the instrument triggers.
REFerence
Outputs a 10 MHz reference frequency.
MASK
Outputs a pulse when a mask is violated. This function is only
available if a mask is specified.
GENerator
Outputs the waveform that is specified using the function gener-
ator (requires option R&S
RTB-B6)
*RST:
OFF
Manual operation:
See
See
TRIGger:OUT:PLENgth
<PulseLength>
Defines the pulse width of the pulse at the [Aux
Out] front connector (at trigger event or
mask viaolation).
Parameters:
<PulseLength>
*RST:
1E-6
Manual operation:
See