Remote control commands
R&S
®
RTB2000
451
User Manual 1333.1611.02 ─ 11
INF
Data points remain on the screen infinitely until persistence is
set to OFF.
*RST:
OFF
Manual operation:
See
DISPlay:PERSistence:TIME
<Time>
Persistence time if persistence is active (
is set to
TIME
.
Each new data point in the diagram area remains on the screen for the duration
defined here. .
Parameters:
<Time>
Persistence time
Range:
50e-3 to 12.8
Increment: 50e-3
*RST:
50e-3
Default unit: s
Manual operation:
See
DISPlay:PERSistence:CLEar
Removes the displayed persistent waveform from the screen.
Usage:
Event
DISPlay:PERSistence[:STATe]
<State>
Defines whether the waveform persists on the screen or whether the screen is
refreshed continuously.
Parameters:
<State>
ON | OFF
ON
The waveform persists for the time defined using
OFF
The waveform does not persist on the screen. Only the currently
measured values are displayed.
*RST:
OFF
General instrument setup