Remote control commands
R&S
®
RTB2000
418
User Manual 1333.1611.02 ─ 11
Example:
DVM:SOUR CH1
DVM:TYPE MEAN
DVM:RES:STAT?
<-- 4.968E-01,5
The result value of the mean measurement on channel 1 is
496.1
mV. The result status is 5 (decimal) = 0101 (binary). That
means, the result is valid (bit 0 = 1), and the signal is clipped by
the limits of the ADC range (bit 3 = 1).
Usage:
Query only
15.8.6
Trigger counter
TCOunter:RESult[:ACTual]:FREQuency?
......................................................................... 418
TCOunter:RESult[:ACTual]:PERiod?
................................................................................ 418
TCOunter:ENABle
<Enable>
Enables or disables the trigger counter measurements.
Parameters:
<Enable>
ON | OFF
TCOunter:SOURce
Sets the measurement source for the counter.
Parameters:
<CounterSource>
CH1 | CH2 | CH3 | CH4 | TRIGger
TRIGger: Trigger source
*RST:
TRIG
TCOunter:RESult[:ACTual]:FREQuency?
Returns the frequency of the trigger source.
Return values:
<FrequencyValue>
Default unit: Hz
Usage:
Query only
TCOunter:RESult[:ACTual]:PERiod?
Returns the period of the trigger source.
Return values:
<PeriodValue>
Default unit: s
Applications