Remote control commands
R&S
®
RTB2000
325
User Manual 1333.1611.02 ─ 11
................................................................................................329
ACQuire:POINts:AUTomatic
<AutoRecordLength>
Enables or disables the automatic record length. The instrument sets a value that fits
to the selected timebase.
If you set a specific value with
ment of a record length is turned off.
Parameters:
<AutoRecordLength> ON | OFF
Example:
ACQ:POIN:AUT ON
TIM:SCAL 1e-9
ACQ:POIN?;:SYST:ERR:ALL?
-> received 10000;0,"No error"
TIM:SCAL 5e-3
ACQ:POIN?;:SYST:ERR:ALL?
-> received 20000000;0,"No error"
Manual operation:
See
See
ACQuire:POINts[:VALue]
Defines a record length value, the number of recorded waveform points in a segment.
The command turns
OFF.
If
ACQuire:POINts?
returns
the automatically set record length.
Each predefined record length corresponds to a maximum number of history seg-
ments, which are stored in the instrument's memory. If option R&S
RTB-K15 is instal-
led, you can display the history segments.
Available record length values are:
●
10 kSa (13107 history segments)
●
20 kSa (13107 history segments)
●
50 kSa (3276 history segments)
●
100 kSa (2621 history segments)
●
200 kSa (1456 history segments)
●
500 kSa (319 history segments)
●
1 MSa (319 history segments)
●
2 MSa (159 history segments)
●
5 MSa (40 history segments)
●
10 MSa (32 history segments)
Waveform setup