Safety Information

R&S

®

RT

Z1M

6

Manual 1801.3142.02 ─ 02

Risk of device damage

The R&S

 

RT

Z1M can withstand a moderate amount of physical and electrical stress.

To avoid damage, treat the adapter with care:

Do not exceed the specified voltage limits.

Connect the R&S

 

RT

Z1M only to an instrument with Rohde

 

&

 

Schwarz probe

interface. Never connect it to a usual BNC jack. Although the 7

 

mm coaxial con-

nector looks like a standard BNC connector, it is constructed differently and does
not fit to the standard BNC jack. The interface of the R&S

 

RT

Z1M can withstand a

higher frequency limit.

Prevent the adapter from receiving mechanical shock.

Do not spill liquids on the adapter.

Store the adapter in a shock-resistant case, e.g. in the foam-lined shipping case.

Содержание R&S RT-Z1M

Страница 1: ...R S RT Z1M 1 MOHM Adapter Manual Manual Version 02 1801314202 B1OZ2...

Страница 2: ...164 Email info rohde schwarz com Internet www rohde schwarz com Subject to change Data without tolerance limits is not binding R S is a registered trademark of Rohde Schwarz GmbH Co KG Trade names ar...

Страница 3: ...ories 10 2 4 2 Service Accessories 10 3 Putting into Operation 11 3 1 Connecting the Adapter to the Oscilloscope 11 3 2 Identification of Adapter and Probe 12 3 3 Connecting the Adapter to the DUT 12...

Страница 4: ...Contents R S RT Z1M 4 Manual 1801 3142 02 02 5 4 Calibration Interval 18 5 5 Discarding the Product 18 6 Functional Check 19...

Страница 5: ...d electric shock and personal injury observe the following instructions The adapter and the measurement instrument must be grounded Grounding is typically ensured through the grounding of the measurem...

Страница 6: ...to an instrument with Rohde Schwarz probe interface Never connect it to a usual BNC jack Although the 7 mm coaxial con nector looks like a standard BNC connector it is constructed differently and does...

Страница 7: ...sive probes current probes and active probes with Rohde Schwarz probe interface The R S RT Z1M provides an integrated R S ProbeMeter which is used for probes without an own R S ProbeMeter Usually thes...

Страница 8: ...completeness and the device has been tested If the packaging material shows any signs of stress notify the carrier and your Rohde Schwarz service center Keep the package and cushioning material for i...

Страница 9: ...The interface provides also a USB interface and supply voltages 5 V USB 12 V high power It transmits analog signals and digital data simultane ously The output of the R S RT Z1M is connected to the os...

Страница 10: ...s Item Quantity Description Order number R S RT ZP10 with probe accessories 1 10 1 passive Probe 1409 7550 00 2 4 2 Service Accessories To order accessories for servicing the probe contact your Rohde...

Страница 11: ...illoscopes that have only inputs with 50 input impedance Oscilloscopes that already support a 1 M input impedance do not support the adapter 3 1 Connecting the Adapter to the Oscilloscope Connect the...

Страница 12: ...Also read and observe the instructions in Chapter 1 Safety Information on page 5 1 Make sure that the oscilloscope is properly grounded 2 Connect the adapter to the oscilloscope 3 Connect the probe to...

Страница 13: ...loscope Figure 3 2 Offset of R S RT Z1M with R S RT ZP10 signal range 1V Div 3 5 2 AC DC Switching Without the adapter the input coupling is limited to DC for a 50 input impedance When the R S RT Z1M...

Страница 14: ...per formed continuously and in parallel to the time domain measurement of the oscillo scope When the R S ProbeMeter is active the measured values are displayed on the oscillo scope The R S ProbeMeter...

Страница 15: ...ransfer functions of the gain or attenuation modes are shown in Figure 4 1 The bandwidth of a system specifies the maximum frequency at which a purely sinus oidal signal is still transferred at 70 3 d...

Страница 16: ...all 7 modes The rise time of a system is inversely proportional to the bandwidth The following approximation applies BW trise 4 0 The minimum measurable rise time of the whole setup also depends on th...

Страница 17: ...rn your R S RT Z1M to your Rohde Schwarz ser vice center A list of all service centers is available on www services rohde schwarz com If you cannot use the original packaging consider the following 1...

Страница 18: ...roduct again make sure to dry it com pletely 5 4 Calibration Interval The recommended calibration interval for R S RT Z1M 1 MOhm adapter is two years For servicing send the probe to your nearest Rohde...

Страница 19: ...esults are influenced by the used oscilloscope Figure 6 1 Functional check 1 Connect the R S RT Z1M to an R S oscilloscope as described in Chapter 3 1 Connecting the Adapter to the Oscilloscope on pag...