
Instrument functions
R&S
®
LCX Series
69
User Manual 1179.2260.02 ─ 02
2. Press the [Trig.] key at the front panel, to start the measurement (Remote com-
The R&S
LCX executes one measurement cycle. When completed, it is waiting for
the next trigger signal.
3. Optionally, configure the next working point.
4. Repeat step
until you have executed the measurement at all work-
ing points.
8.2
Display mode
The R&S
LCX indicates the measurement results numerically and graphically.
Based on the measured values, the instrument calculates and derives the non-measur-
able parameters for display.
8.2.1
Display mode settings
Access:
1. To select the measurement result display, press the [home] key at the front panel.
The numeric display shows the readings of four-measurement pairs by default.
Figure 8-1: Four measurement pairs display
2. Press the [Display Mode] key.
3. Pressing the key repeatedly, toggles between this view, the view with one mea-
surement pair and the graphical view.
Note:
You can also access the graphical display in the "Functions" tab, see
ter 8.8, "Specific instrument functions"
The results display shows the measurement results of the currently selected
parameters. The softkeys enable you to select the parameters for display accord-
ing to your measurement.
The required remote commands are described in:
Display mode
www.allice.de
Allice Messtechnik GmbH