Remote control commands
R&S
®
LCX Series
178
User Manual 1179.2260.02 ─ 02
MAX | MAXimum
Queries the upper limit of the signal level.
DEF | DEFault
Queries the signal level the instrument sets by default.
Range:
0.01 V to 10 V
*RST:
1 V
Default unit: V
Example:
See
Example "Setting key parameters"
Manual operation:
See
11.5
BIAS subsystem
The
BIAS
subsystem contains the commands for bias configuration.
Example: Bias settings
The example represents the remote control commands for applying either an external
or an internal DC bias level.
// Assign an internally generated voltage level and enable bias.
BIAS:VOLTage[:LEVel] 1
BIAS:CURRent[:LEVel] 0.1
BIAS:STATe 1
// Query an externally applied voltage level and enable bias.
BIAS:EXTernal:MEASure:VOLTage?
// Response: 0.2
BIAS:EXTernal[:VOLTage][:STATe]ON
// Response: 1
BIAS:EXTernal:MEASure:VOLTage?
................................................................................ 179
BIAS:EXTernal[:VOLTage][:STATe]
...................................................................................179
BIAS:CURRent[:LEVel]
<Bias Current Level>
BIAS:CURRent[:LEVel]?
<Bias Current Level>
Sets the internal bias current value.
To activate the bias, use command
Parameters for setting and query:
BiasCurrent
Range:
0 A to 0.2 A
*RST:
0 A
Default unit: A
Example:
See
Manual operation:
See
"Current Bias: / [Bias Level]"
BIAS subsystem