29. (V4.11)  GPIB: Basic remote control of the signal generator which is connected to the additional 

FSP-B10 GPIB Interface.  

30. (V4.11)  GPIB: SCPI format for binary block data extended for byte counts > 999.999.999 bytes.  

31. (V4.11)  GPIB: New commands available 

 

:[SENSe<1|2>:]CORRection:TRANsducer:ACTive? 

returns active transducer

 

:CALCulate<1|2>:LIMit<1...8>:ACTive?   

 

returns active limit line(s) 

32. (V4.11)  Trigger Line for video trigger is now also visible outside of the trigger menu.  

33. (V4.11)  Extended resolution for the number of sweep points.  

In addition to currently allowed values an increment of 100 is possible now for number of points 

_

201.

 

34. (V4.11)  HP emulation: The OL command returns the mixer level in byte 23 

35. (V4.11)  HP emulation: 

The commands MKPK NH | NL | NR and KSK do not perform a sweep start

when marker is already switched on

36. (V4.11)  HP emulation: 

The commands SNGLS and CONTS are setting the command complete bit

(bit 4) in STB

37. (V4.11)  HP emulation: New s

oftkey SETUP - GENERAL SETUP - GPIB - SWEEP REP ON/OFF"

38. (V4.11)  HP emulation: 

New commands: VARDEF, CTA, ADD, SUB, MPY, DIV

39. (V4.11)  HP emulation: 

New command NORMLIZE for tracking generator

40. (V4.11)  HP emulation: 

The command LF performs a reset

41. (V4.11)  LXI Class C support 

42. (V4.11)  Measurement  speed  improved  for  EMI  filter  available  (spectrum  mode  /  frequency 

domain).  

43. (V4.11 SP1) New model R&S FSU67 supported.  

44. (V4.11SP1) New CPU Board 1091.3104 supported (with the exception of option B73).  

45. (V4.11SP2) Option B73 now supported with new CPU Board 1091.3104, too.  

46. (V4.21)  New Save/Recall menu and dialogs available.  

47. (V4.21)  An easy access to the Windows XP Start menu is available.  

48. (V4.21)  The required sweeptime is reduced for video bandwidth < resolution bandwidth.  

49. (V4.21)  ASCII Export available for Marker Peak List.  

50. (V4.21)  Adjustable marker position knob stepsize is available.  

51. (V4.21)  New trace average function Power added.  

52. (V4.21)  HP emulation: Personality Spurious supported.  

53. (V4.21)  HP emulation: Personality Phase Noise supported.  

54. (V4.21)  New Service function 

0.6.0 to check installation of option FSU-K51

The function returns 0 / 1 if FSU-K51 is not installed / installed. 
Example: "DIAG:SERV:SFUN? '0.6.0' " 

55. (V4.21)  FSP-B10: Upper frequency limit of SMF100A is now 43.5 GHz .  

56. (V4.21)  The Aquisition Time (for FFT filter) is now readable with remote command 

"SENS:SWE:TIME?".  

57. (V4.21 SP1) FSP-B10: Support for SMA100, SMB100 (1/2/3/6GHz), SMF (22/43GHz) SMJ (3/6GHz).  

58. (V4.21 SP1) FSP-B10: Support for SMF100a - TTL mode.  

59. (V4.21 SP1) FSU-B21 with Order Number 1157.1126.03

 

supported

.

Содержание R&S FSU

Страница 1: ...supports save recall of user standards SEM measurement Supports for save recall of user defined standards SEM measurement Ref Level dialog available to adjust the sweep list s level settings SEM measu...

Страница 2: ...inute changes to the operating manual 16 Manual Operation 16 Quick Start Guide Login 16 Quick Start Guide Operating System Properties Special Links 16 Menu SETUP GENERAL SETUP NEXT 17 Remote Control D...

Страница 3: ...January 2011 5 New functions and improvements with Service Pack 2 added Known issue concerning FS K8 added 09 February 2011 6 Extended remote command description for TRACE LIST for SEM Measurement 11...

Страница 4: ...system and all applications using the new update manager Base System Update from version 4 11 to 4 11 or newer Skip this step if the installed base system firmware is V4 11 or newer The firmware upda...

Страница 5: ...e Switch the instrument off by pressing the ON standby switch at the front panel Switch the power off at the rear panel Wait until the Standby LED on the front panel turns from yellow to black off Swi...

Страница 6: ...ng the license key for every option is SETUP GENERAL SETUP OPTIONS INSTALL OPTION Use the numeric keypad to input the option key number and press ENTER On a successful validation the message option ke...

Страница 7: ...Bits MEASuring SWEeping with Service Pack 2 Improvements The version numbers in brackets indicate the version in which the issue was observed for the first time 1 V4 41 The remote command MMEM DATA ig...

Страница 8: ...ep Points is not set The follow configuration files for EUTRA LTE Uplink needs 30001 sweep points to be set BW_01_4_MHz xml BW_03_0_MHz xml BW_05_0_MHz xml The number of sweep points is no automatical...

Страница 9: ...NERAL SETUP OPTIONS A reboot is required for a few options The message box The system must be rebooted to effect the changes Reboot now will be indicated in that case With versions 4 61 4 61 SP1 the s...

Страница 10: ...affected Work around Switch back to analyzer before activation of the next application Example RST SYST DISP UPD ON INST SEL BWCD activate K72 do some measurements in K72 INST SEL SAN back to analyzer...

Страница 11: ...odels 71100C 71200C and 71209A are using 800 sweep points 13 V4 01 Additional number of sweep points 201 401 801 and 1601 14 V4 01 HP emulation Additional HP Models 8568A_DC and 8568B_DC using DC coup...

Страница 12: ...0 V4 11 HP emulation The command LF performs a reset 41 V4 11 LXI Class C support 42 V4 11 Measurement speed improved for EMI filter available spectrum mode frequency domain 43 V4 11 SP1 New model R S...

Страница 13: ...69 V4 31 Dummy Video Bandwidt 0 Hz returned for active FFT filter 70 V4 31 Availability changed for Spurious Measurement The Spurious Measurement is not available if the ACP measurement is active 71 V...

Страница 14: ...ase Noise Measurement some minor adjustments Support of 4 markers Corrections of RBW calculation if FFT Filter is switched on Command MKNOISE MKTRACK correction of return value Sweep time adjusted for...

Страница 15: ...EM Measurement Required Number of Sweep Points is not set The follow configuration for EUTRA LTE Uplink needs 30001 sweep points to be set BW_01_4_MHz xml BW_03_0_MHz xml BW_05_0_MHz xml The number of...

Страница 16: ...by entering a user name and password in a login window The instrument provides a factory installed auto login function i e login is carried out automatically in the background The ID used for auto log...

Страница 17: ...3 More details see chapter SETUP GENERAL SETUP Menu SETUP GENERAL SETUP NEXT REGISTRY READ ONLY The softkey REGISTRY READ ONLY activates deactivates a write protection for the Windows XP registry Any...

Страница 18: ...ming a calibration 1 to 2 These bits are not used 3 SWEeping This bit is set while the instrument performs a sweep It is supported in analyzer mode only Full Screen frequency domain and time domain 4...

Страница 19: ...a the SENSe 1 2 LIST RANGe 1 20 subsystem spurious measurement or the SENSe 1 2 ESPectrum RANGe 1 20 subsystem spectrum emission mask measurement SPURious reads the peak list in the spurious measureme...

Страница 20: ...Supported 89600 commands Status CAL available in V4 61 SP2 and above CLS available in V4 61 SP2 and above ESE available in V4 61 SP2 and above ESR available in V4 61 SP2 and above IDN available in V4...

Страница 21: ...RLEVel numeric available in V4 61 SP2 and above DISPlay WINDow TRACe Y SCALe RLEVel OFFSet numeric available in V4 61 SP2 and above FORMat BORDer NORMal SWAPped available in V4 61 SP2 and above FORMat...

Страница 22: ...l FREQuency numeric available in V4 61 SP2 and above SENSe ROSCillator OUTPut OFF ON available in V4 61 SP2 and above SENSe ROSCillator SOURce INTernal EXTernal EAUTo available in V4 61 SP2 and above...

Страница 23: ...ble in V4 61 SP2 and above STATus QUEStionable CALibration NTRansition number available in V4 61 SP2 and above STATus QUEStionable CALibration PTRansition number available in V4 61 SP2 and above STATu...

Страница 24: ...hour minute second available in V4 61 SP2 and above SYSTem VERSion available in V4 61 SP2 and above TRACe COPY src_trace dest_trace available in V4 61 SP2 and above TRACe DATA TRACE1 TRACE2 TRACE3 TR...

Страница 25: ...ger SEQuence SOURce IMMediate VIDeo EXTernal 1 2 available in V4 61 SP2 and above TRIGger SEQuence VIDeo LEVel numeric available in V4 61 SP2 and above TRIGger SEQuence VIDeo LEVel FREQuency freq avai...

Страница 26: ...AR RANGE LINEAR dB FM FM AF Spektrum PM PM AF Spektrum AM AM AF Spektrum RF Power RF Power AF Spektrum RF Spektrum RF POWER RF SPECTRUM Result Display DIAGRAM FULL SIZE AM AF SPECTRUM ZERO PHASE REF P...

Страница 27: ...llowing demodulation bandwidth range 100 kHz demodulation bandwidth 3 MHz A WEIGHTED Switches on the A weighted filter The weighting filter is active in the following demodulation bandwidth range 100...

Страница 28: ...calculations The query command is available only with CALC ADEM THD FREQ FUND AUTO OFF Example CALC ADEM THD FREQ FUND AUTO OFF deactivates the auto se lection and uses the current Modulation Freq as...

Страница 29: ...ghted CCIR filter Characteristics RST Wert OFF SCPI device specific The numeric suffix 1 2 is irrelevant with this command R S FS K8 Extensions The additional Enhanced Data Rate functions are describe...

Страница 30: ...r calculates the signal pulse power from this value and the mean power The softkey is highlighted if the correction is switched on Press the softkey again to switch the Duty Cycle correction off Valid...

Страница 31: ...al entries is 0 001 The default setting is 99 999 Example SENS PMET STAT ON activate power meter SENS PMET DCYC STAT ON switch the correction on SENS PMET DCYC VAL 50 0 set the duty cycle to 50 Proper...

Страница 32: ...p to date information and upgrades To keep your instrument up to date and to be informed about new application notes related to your instrument please send an e mail to the Customer Support Center sta...

Отзывы: