Remote Commands
R&S
®
Spectrum Rider FPH
396
User Manual 1321.1011.02 ─ 06.00
If you use an external reference signal, make sure to connect the signal to the Ext Ref
BNC connector of the R&S Spectrum Rider.
Parameters:
<Format>
INTernal | EXTernal
INTernal
Internal reference.
EXTernal
External reference.
Example:
ROSC:SOUR EXT
Activates external source as reference signal.
SYSTem:ACCessory
<Accessory>
This command queries the type of measurement accessory, if one is connected to the
R&S Spectrum Rider (for example a power sensor).
Parameters:
<Accessory>
Z1 | Z2 | Z3 | Z14 | Z18 | Z44 | TS_emf | NONE | UNKNown
Name of the accessory.
Example:
SYST:ACC?
Queries connected measurement accessories.
SYSTem:ACCessory:AUTO
<State>
This command turns automatic detection of connected measurement accessories on
and off.
Parameters:
<State>
ON | OFF
Example:
SYST:ACC:AUTO ON
Turns on automatic accessory detection.
SYSTem:BNC:MODE
<BNCUsage>
This command configures the BNC sockets.
Parameters:
<BNCUsage>
REFerence | TRIGger | BIAS
REFerence
Input for external reference signal.
TRIGger
Input for external trigger.
BIAS
BIAS port.
*RST:
TRIGger
Remote Control - Commands