Remote Commands
R&S
®
Spectrum Rider FPH
298
User Manual 1321.1011.02 ─ 06.00
Parameters:
<Unit>
DBM | DBMV | DBUV | VOLT | WATT | DUVM | DUAM | V | W |
V_M | W_M2 | RHO | MRHO
Note that the availability of units depends on the operating
mode.
*RST:
DBM
Example:
UNIT:POW DBUV
Sets the power unit to dBμV.
13.6.2.3
Setting the Bandwidths
The following commands configure the filter bandwidths of the R&S Spectrum Rider.
Note that both groups of commands (
BANDwidth
and
BWIDth
) are the same.
List of commands
●
[SENSe:]BANDwidth[:RESolution]
●
●
[SENSe:]BANDwidth[:RESolution]:AUTO
●
[SENSe:]BWIDth[:RESolution]:AUTO
●
●
●
●
[SENSe:]BANDwidth[:RESolution]
<ResolutionBW>
This command defines the resolution bandwidth.
Parameters:
<ResolutionBW>
Range:
1 Hz to 3 MHz
*RST:
- (AUTO is set to ON)
Default unit: GHz
Example:
BAND 100 kHz
Sets the resolution bandwidth to 100 kHz.
[SENSe:]BWIDth[:RESolution]
<ResolutionBW>
This command defines the resolution bandwidth.
Parameters:
<ResolutionBW>
Range:
1 Hz to 3 MHz
*RST:
- (AUTO is set to ON)
Default unit: GHz
Example:
BWID 100 kHz
Sets the resolution bandwidth to 100 kHz.
Remote Control - Commands