NB-IoT Multi-Evaluation Measurement
R&S
®
CMW-KM300/-KS300
281
User Manual 1178.5689.02 ─ 06
CONFigure:NIOT:MEAS<i>:MEValuation:LIST:OSINdex
<OfflineSegIndex>
Selects the segment to be displayed in offline mode. The index refers to the range of
measured segments, see
CONFigure:NIOT:MEAS<i>:MEValuation:LIST:
.
Setting a value also enables the offline mode.
Parameters:
<OfflineSegIndex>
Range:
1 to <number of measured segments>
*RST:
OFF
Additional parameters: OFF (disables the offline mode)
Example:
See
Performing Single-Shot Measurements
Firmware/Software:
V3.7.20
Options:
R&S CMW-KM012
Manual operation:
See
"Offline Segment Nr. < List Mode"
CONFigure:NIOT:MEAS<i>:MEValuation:LIST:CMWS:CMODe
<ConnectorMode>
Specifies how the input connector is selected for NB-IoT list mode measurements with
the R&S
CMWS.
Parameters:
<ConnectorMode>
GLOBal | LIST
GLOBal:
The same input connector is used for all segments. It
is selected in the same way as without list mode, for example via
ROUTe:NIOT:MEAS<i>:SCENario:SALone
.
LIST:
The input connector is configured individually for each
segment. See
CONFigure:NIOT:MEAS<i>:MEValuation:
LIST:SEGMent<no>:CMWS:CONNector
*RST:
GLOB
Example:
See
Firmware/Software:
V3.7.20
Options:
R&S CMW-KM012
CONFigure:NIOT:MEAS<i>:MEValuation:LIST:SEGMent<no>:CMWS:CONNector
<CMWSConnector>
Selects the RF input connector for segment <no> for NB-IoT list mode measurements
with the R&S
CMWS. This setting is only relevant for connector mode
LIST
, see
CONFigure:NIOT:MEAS<i>:MEValuation:LIST:CMWS:CMODe
.
All segments of a list mode measurement must use connectors of the same bench.
For possible connector values, see
Chapter 3.5.1.4, "Values for RF Path Selection"
Command Reference