NB-IoT Multi-Evaluation Measurement
R&S
®
CMW-KM300/-KS300
278
User Manual 1178.5689.02 ─ 06
Parameters:
<Lagging>
OFF | S1
OFF
: no exclusion
S1
: exclude first symbol
*RST:
OFF
Example:
See
Specifying Measurement-Specific Settings
Firmware/Software:
V3.7.30
Manual operation:
See
"EVM Excl. Periods Leading/Lagging < Modulation"
CONFigure:NIOT:MEAS<i>:MEValuation:SCOunt:MODulation
<StatisticCount>
Specifies the statistic count of the measurement. The statistic count is equal to the
number of measurement intervals per single shot.
Parameters:
<StatisticCount>
Number of measurement intervals (slots)
Range:
1 slot to 1000 slots
*RST:
20 slots
Example:
See
Specifying Measurement-Specific Settings
Firmware/Software:
V3.5.10
Manual operation:
See
CONFigure:NIOT:MEAS<i>:MEValuation:SCOunt:SPECtrum:SEMask
<StatisticCount>
Specifies the statistic count of the measurement. The statistic count is equal to the
number of measurement intervals per single shot.
Parameters:
<StatisticCount>
Number of measurement intervals (slots)
Range:
1 slot to 1000 slots
*RST:
20 slots
Example:
See
Specifying Measurement-Specific Settings
Firmware/Software:
V3.5.10
Manual operation:
See
CONFigure:NIOT:MEAS<i>:MEValuation:SCOunt:SPECtrum:ACLR
<StatisticCount>
Specifies the statistic count of the measurement. The statistic count is equal to the
number of measurement intervals per single shot.
Command Reference