Remote control commands
R&S
®
NRPM
184
Mannual 1425.8663.02 ─ 08
10.11.6
Controlling the positive transition part
For more information on the positive transition part see
STATus:DEVice:PTRansition
<value>
STATus:OPERation:CALibrating:PTRansition
<value>
STATus:OPERation:PTRansition
<value>
STATus:OPERation:LLFail:PTRansition
<value>
STATus:OPERation:MEASuring:PTRansition
<value>
STATus:OPERation:SENSe:PTRansition
<value>
STATus:OPERation:TRIGger:PTRansition
<value>
STATus:OPERation:ULFail:PTRansition
<value>
STATus:QUEStionable:CALibration:PTRansition
<value>
STATus:QUEStionable:PTRansition
<value>
STATus:QUEStionable:POWer:PTRansition
<value>
These commands control the Positive TRansition part of a register.
Setting a bit causes a 0 to 1 transition in the corresponding bit of the associated
register. The transition also writes a 1 into the associated bit of the corresponding
EVENt
register.
Parameters:
<value>
*RST:
65535
10.12
Testing the R&S
NRPM OTA Power Measure-
ment Solution
The selftest allows a test of the internal circuitry of the sensor.
TEST:SENSor?
[<Item>]
Triggers a selftest of the sensor module. In contrast to
returns detailed information that you can use for troubleshooting.
Note:
Do not apply a signal while the selftest is running because it can indicate
errors in the output messages for the test steps
Offset Voltages
and/or
Noise Vol-
tages
.
Testing the R&S
NRPM OTA Power Measurement Solution