SYSTem:COMMunicate:NETWork:IPADdress
<ipaddress>
R&S
NRPM3N LAN sensor modules only
SYSTem:COMMunicate:NETWork:IPADdress:MODE STATic
set the address manually.
Sets the IP address.
Parameters:
<ipaddress>
Range:
0.0.0.0. to ff.ff.ff.ff
Example:
SYSTem:COMMunicate:NETWork:IPADdress
"10.113.0.104"
Manual operation:
SYSTem:COMMunicate:NETWork:IPADdress:GATeway
<gateway>
R&S
NRPM3N LAN sensor modules only
Sets the IP address of the default gateway.
SYSTem:COMMunicate:NETWork:IPADdress:MODE STATic
set the gateway manually.
Parameters:
<gateway>
Example:
SYSTem:COMMumicate:NETWork:IPADdress:
GATeway '192.168.10.254'
Sets 192.168.10.254 as IP address of the default gateway.