Preparing for Use
R&S
®
NRP Series
13
Getting Started 1419.0170.02 ─ 12
Tighten the RF connector manually.
4. To ensure maximum measurement accuracy, tighten the RF connector using a
torque wrench with the nominal torque recommended in
To disconnect from the DUT
►
NOTICE!
Risk of damaging the center pin of the RF connector. Always rotate
only the hex nut of the RF connector. Never rotate the power sensor itself.
Carefully loosen the union nut at the front of the RF connector of the sensor
and remove the sensor.
4.5
Connecting a Cable to the Host Interface
For connecting the power sensor to a USB host, use the host interface. See
Depending on the USB host, use one of the following cables:
●
Computer or R&S NRP
‑
Z5 sensor hub:
R&S NRP
‑
ZKU cable with a USB connector, R&S order number 1419.0658.xx
●
Base units, R&S NRX or R&S NRP2, or other supported Rohde
&
Schwarz
instruments:
R&S NRP
‑
ZK6 cable with a push-pull type connector, R&S order number
1419.0664.xx
See
These cables can be obtained in different lengths up to 5 meters.
To connect a cable to the host interface of the power sensor
1. Insert the screw-lock cable connector into the host interface connector of the
power sensor.
2. Tighten the union nut manually.
To disconnect the host interface of the power sensor
► Loosen the union nut of the screw-lock cable connector and remove the cable.
Connecting a Cable to the Host Interface