Remote Control Commands
R&S
®
NRPxxTWG
73
User Manual 1178.8371.02 ─ 04
Performing Configuration Tasks
In this chapter, different configuration tasks performed with the power sensor and the
"S-Parameter Update Multi" tool are described.
To load a calibration data set from a power sensor
Prerequsites: The power sensor is connected to the computer and a connection is
established.
1. Open the "S-Parameter Update Multi" program.
2. Select "Sensor" > "Load Calibration Data".
The "Upload Calibration Data" dialog opens. It shows a list of the available sen-
sors.
3. If you cannot find your power sensor in the list, for example because of reconnect-
ing the power sensor, you can reload the list by clicking "Rebuild List".
4. Click "Upload" to load calibration data from the power sensor.
After the upload is finished, the "OK" button is enabled.
5. Click "OK" to apply the changes.
If you want to discard the changes, exit the dialog by clicking "Cancel".
After a successful upload, the name and serial number is shown in the name of the
main dialog.
Configuring Basic Measurement Parameters