Remote Control Commands
R&S
®
NRPxxTWG
60
User Manual 1178.8371.02 ─ 04
[SENSe<Sensor>:][POWer:][AVG:]SMOothing:STATe
<state>
Enables or disables the smoothing filter, a steep-edge digital lowpass filter. If you can-
not adjust the aperture time exactly to the modulation period, the filter reduces result
fluctuations caused by modulation.
Parameters:
<state>
ON | OFF
*RST:
OFF
Example:
SMO:STAT OFF
7.8
Configuring Basic Measurement Parameters
The following section describes the settings common for several measurement modes.
7.8.1
Configuring Auto Averaging
Describes the commands for automatic averaging in the continuous average measure-
ments.
Remote commands:
..................................................................................60
[SENSe<Sensor>:]AVERage:COUNt:AUTO
........................................................................61
[SENSe<Sensor>:]AVERage:COUNt:AUTO:MTIMe
.............................................................61
[SENSe<Sensor>:]AVERage:COUNt:AUTO:NSRatio
...........................................................61
[SENSe<Sensor>:]AVERage:COUNt:AUTO:RESolution
.......................................................62
...................................................................................62
[SENSe<Sensor>:]AVERage:COUNt:AUTO:TYPE
.............................................................. 62
[SENSe<Sensor>:]AVERage:TCONtrol
.............................................................................. 63
[SENSe<Sensor>:]AVERage[:STATe]
................................................................................. 63
[SENSe<Sensor>:]AVERage:COUNt
<count>
Sets the number of readings that are averaged for one measured value. The higher the
count, the lower the noise, and the longer it takes to obtain a measured value.
Average count is often also called averaging factor, but it designates the same thing,
i.e the number of measured values that have to be averaged for forming the measure-
ment result.
Averaging is only effective, if
[SENSe<Sensor>:]AVERage[:STATe]
ON
is set.
Parameters:
<count>
Range:
1 to 65536
*RST:
4
Example:
AVER:COUN 1
Configuring Basic Measurement Parameters