Instrument functions
R&S
®
NGP800
53
User Manual 5601.5610.02 ─ 09
Slew rate control
●
The slew rate control does not affect the output turn-on slew rate programmed via
the
●
The slew rate control mode is not compatible with
tions. The reduce slew rate function is automatically disable if any of these func-
tions is used and vice versa.
●
The minimum setting voltage is 0.003 V if "Reduce Slew Rate" is activated, e.g.
output voltage is automatically set to 0.003 V if setting value is < 0.003 V.
► Select "Reduce Slew Rate" menu item to activate the slew rate control.
The R&S
NGP800 activates the slew rate control.
6.3
Analog input
Instrument option
R&S NGP-K107 (P/N: 5601.6200.03) option is required for the "Analog Input".
Analog input
If "Analog Input" of a selected channel (voltage or current) is enabled, the respective
channel voltage or current setting is disabled.
See
Chapter 6.1, "Setting the channels voltage and current"
With "Analog Input", you can regulate the channel output (voltage or current) settings
with an input voltage of 0 V to 5 V.
Analog input mode
NGP output for 64 V channels
NGP output for 32 V chan-
nels
Voltage mode
0 V to 64 V
0 V to 32 V
Current mode
0 A to 10 A
0 A to 20 A
1. Press [Settings] key.
The R&S
NGP800 displays the device/channel menu window.
2. Select the required channel tab to configure the analog input mode.
The R&S
NGP800 displays the selected channel menu.
3. Select "Analog Input" from the menu.
The R&S
NGP800 displays the selected "Analog Input" dialog.
Analog input