2

Technical Data

Max. input voltage

100 V

(derates at 20 db/decade to 5 V above 

100 kHz)

Trigger coupling

AC

  R&S

®

HMO1002 series

  R&S

®

HMO1202 series

10 Hz  to  50/70/100 MHz
10 Hz  to  100/200/300 MHz

DC

  R&S

®

HMO1002 series

  R&S

®

HMO1202 series

DC to 50/70/100 MHz
DC to 100/200/300 MHz

Trigger Output via AUX OUT (BNC)

 

Functions

Pulse output for every acquisition trigger 
event, error output on mask violation

 

Output level

approx.  3 V

 

Pulse polarity

positive

 

Pulse width

>150 ns (trigger event),
>0.5 µs (mask violation)

Trigger Types
Edge

Direction

rising, falling, both

Trigger coupling

auto level AC, DC, HF

Switchable filters

LF, noise rejection

Sources

 R&S®HMO1002, 
 R&S®HMO1202

all analog and digital channels, AC line, 
external (AC, DC)

Pulse Width

Polarity

positive, negative

Functions

equal, not equal, lower, higher, within/without 
a range

Pulse duration

16ns to 10s, resolution min. 2ns

Sources

all analog channels

Logic 

Functions
  boolean operators

AND, OR, TRUE, FALSE

  time based operators

equal, not equal, lower, higher, within/without 
a time range, timeout

Duration

16 ns to 10 s, resolution min. 2 ns

States

H, L, X

Sources

all logic channels

Video

Sync. pulse polarity

positive, negative

Supported standards

NTSC, SECAM, PAL, PAL-M, SDTV 576i, 
HDTV 720p, HDTV 1080i, HDTV 1080p

Field

even/odd, either

Line

line number selectable, all

Sources

all analog channels, external (AC, DC)

Serial Busses (optional)

  Bus representation

Up to two busses can be analyzed at the 
same time. Color-coded display of decoded 
data in ASCII, binary, decimal and 
hexadecimal format.

Option / Voucher codes

  R&S®HOO10, R&S®HV110 Analysis of I

2

C, SPI, UART/RS-232 signals on 

analog and logic channels

  R&S®HOO11, R&S®HV111 Analysis of I

2

C, SPI, UART/RS-232 signals on 

all analog channels

  R&S®HOO12, R&S®HV112 Analysis of CAN and LIN signals on analog 

and logic channels

Trigger types by protocols
 I

2

C

Start, Stop, ACK, NACK, Address/Data

 SPI

Start, End, Serial Pattern (32 Bit)

 UART/RS-232

Startbit, Frame Start, Symbol, Pattern

 LIN

Frame Start, Wake Up, Identifier, Data, Error

 CAN

Frame Start, Frame End, Identifier, Data, Error

Horizontal System

Time domain (Yt)

main screen, time domain and zoom window

Frequency domain (FFT)

time domain and frequency domain window 
(FFT)

XY mode

voltage (XY)

VirtualScreen

virtual display of ±10 Div for all math, logic, 
bus, reference signals

Component tester

voltage (X), current (Y)

Reference signals

up to 4 references

Channel deskew

±32 ns, step size 2 ns

Memory zoom

up  to  50.000 : 1

Time base
 accuracy

±50.0 x 10

-6

 aging

±10.0 x 10

-6

 per year

Operation modes
 REFRESH

  R&S

®

HMO1002 series

  R&S

®

HMO1202 series

2 ns/Div  to  50 s/Div
1 ns/Div  to  50 s/Div

 ROLL

50 ms/Div  to  50 s/Div

Acquisition System
Realtime Sampling Rate

Analog channels

 R&S

®

HMO1002 series

 R&S

®

HMO1202 series

2x 500 MSa/s or 1x 1 GSa/s
2x 1 GSa/s or 1x 2 GSa/s

Logic channels

 R&S

®

HMO1002 series

 R&S

®

HMO1202 series

8x  500 MSa/s
8x  1 GSa/s

Memory depth

Analog channels

 R&S

®

HMO1002 series

 R&S

®

HMO1202 series

2x 500 kSa or 1x 1 MSa
2x 1 MSa or 1x 2 MSa

Logic channels
 R&S

®

HMO1002 series

 R&S

®

HMO1202 series

500 kSa per channel
1 MSa per channel

Resolution

8 Bit, (HiRes up to 16Bit)

Waveform arithmetics

refresh, roll (loose/triggered), average (up to 
1024), envelope, peak detect (2 ns), filter (low-

pass, adjustable), high resolution (up to 16 bit)

Record modes

automatic, max. sampling rate,
max. waveform rate

Interpolation

  all analog channels

sin(x)/x, linear, sample-hold

  logic channels

pulse

Delay

 pre-trigger

  R&S

®

HMO1002 series

  R&S

®

HMO1202 series

0 to 500.000 Sa x (1/sample rate)
0 to 1.000.000 Sa x (1/sample rate)
(multiplied by 2 in interlaced mode)

 post-trigger

0 to 8x10

6

 Sa x (1/sample rate)

Waveform update rate

up to 10,000 Wfm/s

3

Technical Data

Waveform display

dots, vectors, persistence afterglow

Persistence afterglow

min.  50 ms

Waveform Measurements and Operation

Operation

menu-driven (multilingual), auto-set, help 
functions (multilingual)

Automatic measurements

voltage (V

pp

, V

p+

, V

p-

, V

rms

, V

avg

, V

min

, V

max

), 

amplitude, phase, frequency, period, rise/fall 
time (80%, 90%), pulse width (pos/neg), 
burst width, duty cycle (pos/neg), standard 
deviation, delay, crest factor, overshoot (pos/
neg), edge/pulse count (pos/neg), trigger 
period, trigger frequency

Cursor measurements

voltage (V1, V2, ∆V), time (t1, t2, ∆t, 1/∆t), 

ratio X, ratio Y, pulse and edge count (pos/
neg), peak values (V

pp

, V

p+

, V

p-

), V

mean

, V

rms

standard deviation, duty cycle (pos/neg), rise/
fall time (80%, 90%), ratio marker, crest factor

Quick measurements
(QUICKVIEW)

voltage (V

pp

, V

p+

, V

p-

, V

rms

, V

mean

), rise/fall 

time, frequency, period plus 6 additional 
measurement functions (see automatic 
measurement functions, freely selectable)

Marker

up to 8 freely positionable markers for easy 
navigation

Frequency Counter (hardware based)

Resolution

5 digit

Frequency range

 R&S

®

HMO1002

 R&S

®

HMO1202

0.5 Hz  bis  50/70/100 MHz
0.5 Hz  bis  100/200/300 MHz

Accuracy

±50.0 x 10

-6

Aging

±10.0 x 10

-6

 per year

Mask Testing

Functions

Pass/Fail comparison with an user-definied 
mask performed on waveforms

Sources

all analog channels

Mask definition

Mask enclosing acquired waveform with user-
defined tolerance

Actions

  on mask violations

beep, acquisition stop, screenshot, trigger 
pulse, automatically saving trace data

  during acquisiton

statistics: number of completed tests, number 
of passes / failed acquisitions (absolute and in 
percent), test duration

Waveform Maths
Quickmath

Functions

addition, substraction, multiplication, division

Sources

CH1, CH2

Mathematics (R&S®HMO1202 series)

Functions

addition, substraction, multiplication, division, 
minimum / maximum, square, square root, 
absolute value, pos/neg wave, reciprocal, 
inverse, log10/ln, derivation, integration, filter 
(lowpass/highpass)

Editing

formula editor, menu-driven

Sources

all analog channels, user-defined constants

Storage location

math. memory

Number of formula sets

5 formula sets

Number of equations

5 equations per formula set

Simultaneous display of 
math. functions

1 formula set with max. 4 equations

Frequency Analysis (FFT)

Parameters

frequency span, center frequency, vertical 
scale, vertical position

FFT length

2 Kpts,  4 Kpts,  8 Kpts,  16 Kpts,  32 Kpts,  64 
Kpts,  128 Kpts

Window

Hanning, Hamming, Rectangular, Blackman

Scale

dBm, dBV, V

rms

Waveform arithmetics

refresh, envelope, average (up to 512)

Cursor measurement

2 horizontal cursors, previous / next peak 
search

Sources

all analog channels

Probe Adjust Output

Operation

manual, adjust-wizzard

Frequence

1 kHz,  1 MHz

Level

 R&S

®

HMO1002 series

 R&S

®

HMO1202 series

approx.  2.5 V

pp

 (t

a

  <4 ns)

approx.  2.5 V

pp

 (t

a

  <1 ns)

Pattern Generator 

Functions

square wave / probe adjust, bus signal source, 

counter, programmable pattern

Square wave
(Probe ADJ output)

frequency range: <1 mHz to 500 kHz
level:  2.5 V

pp

 (ta <4 ns)

polarity: normal, invert
duty cycle: 1% to 99%

Bus Signal Source (4 Bit)

I

2

C (100 kBit/s, 400 kBit/s, 1 MBit/s), SPI 

(100 kBit/s,  250 kBit/s,  1 MBit/s),  UART 
(9600 Bit/s,  115,2 kBit/s,  1 MBit/s), 
CAN (up to 50 MBits/s), 
LIN (up to 50 MBits/s)

Counter (4 Bit)

frequency: <1 mHz to 25 MHz
direction: incrementing, decrementing

Programmable pattern 
(4 Bit)

sampling time: 20 ns to 42 s
memory depth: 2048 sa
pattern idle time: 20 ns to 42 s

Function Generator

Waveform modes

DC, sine, square, triangle/ramp, pulse

 Sine

frequency range: 0.1 Hz to 50 kHz
flatness: ±1 dB relative to 1 kHz
DC offset: max. ±3 V

 Square

frequency range: 0.1 Hz to 25 kHz
rise time: <4 µs
DC offset: max. ±3 V

  Triangle / Ramp

frequency range: 0,1 Hz bis 10 kHz
DC offset: max. ±3 V

 Pulse

frequency range: 0.1 Hz to 10 kHz
duty cycle: 10% to 90%
DC offset: max. ±3 V

Sampling rate

978 kSa/s

Frequency accuracy

±50.0 x 10

-6

Aging

±10.0 x 10

-6

 per year

Amplitude
 DC

±3 V

  DC offset error (meas.)

±25 mV  (max.)

  high impedance load

60 mV

pp

 to 6 V

pp

  50 Ω load

30 mV

pp

 to 3 V

pp

 accuracy

3%

Digital Voltmeter

Display (3-digit)

Primary and secondary measurement value 
per channel, simultaneous measuring on all 
channels

Functions

DC, DC

rms

, AC

rms

, V

pp

, V

p+

, V

p

-, crest factor

Sources

all analog channels

Содержание HMO1002 Series

Страница 1: ...Test Measurement User Manual Version 04 R S HMO1002 Series R S HMO1202 Series Digital Oscilloscope User Manual 5800530103...

Страница 2: ...shorter maximum cable length is prescribed signal cables input output signal control may not exceed a length of 1m and may not be used outside of buildings All signal cables must be shielded coaxial c...

Страница 3: ...Function 21 5 7 Marker Function 21 6 Trigger System 22 6 1 Trigger Modes Auto Normal and Single 22 6 2 Trigger Source 22 6 3 Trigger type 22 6 4 Trigger Events 25 6 5 External Trigger 25 7 Signal Disp...

Страница 4: ...nce with the regulations of protection class I For safety reasons the instrument may only be operated with authorized safety sockets The power cord must be plugged in before signal circuits may be con...

Страница 5: ...be carried out only by authorized ROHDE SCHWARZ technical personnel Only original parts may be used for replacing parts relevant to safety e g power swit ches power transformers fuses A safety test m...

Страница 6: ...served either at all or to the extent necessary product users may be exposed to the risk of explosions fire and or serious personal injury and in some cases death Batteries and rechargeable batteries...

Страница 7: ...rical and Electronic Equipment Act ElektroG ROHDE SCHWARZ assumes full responsibil ity for the ecological disposal or the recycling at the end of life of their products Please contact your local servi...

Страница 8: ...y for the logic probes of type R SHO3508 Connecting other types may demolish the input Section A This section includes the CURSOR MENU ANALYZE and GENERAL sections The CURSOR MENU section includes cur...

Страница 9: ...lly de rived curves are shown in the lower section of the screen Within the graticule signals of the selected channels are displayed By default 8 scale divisions are shown This can be extended virtual...

Страница 10: ...ent Firmware Update The R S HMO1002 resp R S HMO1202 series is con stantly being advanced You can download the current firmware at www rohde schwarz com The firmware is packed in a ZIP file After down...

Страница 11: ...rm the entry The option will be activated after restarting the instrument The R S HMO1002 resp R S HMO1202 series is prepa red for mixed signal operation and have the appropriate connectors on the fro...

Страница 12: ...rked in blue a Fig 2 7 Self alignment logic probe message will display on the screen In addition the home screen the instrument information and a screenshot will indicate that the education mode has b...

Страница 13: ...ect frequently used settings Press the top soft menu key to change the input coupling to Make sure to align the passive probes prior to their first use For more information please refer to the probe m...

Страница 14: ...easurements of the signal Again briefly press the AUTOSET key once followed by the CURSOR MEASURE key You can now select the desired type of measurement cursor from the menu Press the soft menu key ME...

Страница 15: ...nnector of your oscilloscope Press the Save Recall 12 key to open the respective menu Select the type of data you wish to store by pressing the res pective soft menu key in this example SCREENSHOTS Th...

Страница 16: ...n channels by pressing the respective channel key 4 2 Sensitivity Y Positioning and Offset You can select the sensitivity of the analog inputs by using the knob in the VERTICAL section VOLTS DIV in 1...

Страница 17: ...ft menu SQUARE WAVE see chapter 10 2 1 4 5 Threshold Setting On page 2 2 in the advanced MENU in the VERTICAL section you can set a THRESHOLD This threshold defines the level to detect a High or a Low...

Страница 18: ...eature is not available in the roll mode also refer to Chap 5 5 ZOOM Function 5 HorizontalSystem In the HORIZONTAL system section users can select time base settings for capturing trigger time positio...

Страница 19: ...detection mode is marked with HR in the upper right of the display The following requirements must be met to activate the HIGH RESOLUTION mode Sampling rate is smaller than the maximum sampling rate n...

Страница 20: ...t of displayed data will be adjusted Peak detection is also activated in STOP mode time base in microse conds In time bases displaying each sampling point all three settings behave identically except...

Страница 21: ...in will highlight the cursors limiting the zoom area in white allowing you to use the knob to change the zoom area Now you can use the small knob in the horizontal area of the control panel to move th...

Страница 22: ...nal that does not fulfill the trigger condition a simple case would be direct current the oscilloscope itself will generate a trigger event This allows a glance at the input signals at any time regard...

Страница 23: ...ble with AC and DC coupling NR noise reduction A low pass filter with a maximum cut off frequency of 100MHz will improve the noise performance for the trigger amplifier This filter removes high freque...

Страница 24: ...logic trigger without any active logic probes R S HO3508 connected Fig 6 3 Logic trigger menu ti t The duration of the applied bit pattern which will generate the trigger is equal to the adjustable r...

Страница 25: ...e the oscilloscope will trigger on the start of the half frames in the video signal The respective key will select the odd even half frames i e even if the other trigger conditions are met the oscillo...

Страница 26: ...the screen depending on the particular application e g values are displayed when offset is changed A trans parency value of 0 to 100 is selectable Use the universal knob or the KEYPAD key to define th...

Страница 27: ...e displayed more brightly than older signals For instance if 300ms is selected the display for the signal curves will become darker in 50ms intervals and will be erased after 300ms The AUTOMATIC setti...

Страница 28: ...measurement values are displayed in four different formats number of rising and falling slopes and number of positive and negative pulses PEAK VALUES This mode provides two cursors to measure the min...

Страница 29: ...section Fol lowing automatic measurement types can be selected Fig 8 1 Cursor measurement setting menu MEAN VALUE This mode measures the mean value of the signal amplitude Mean If the signal is perio...

Страница 30: ...e screen tr The rise time identifies the time in which the signal rises from 10 to 90 of its amplitude FALL TIME 90 This mode measures the fall time of the first falling slope in the displayed view of...

Страница 31: ...layed view of the screen The mean value is calculated from the amplitude of the measurement signal A slope will be counted if the signal runs through the mean value The measurement will only be applie...

Страница 32: ...e following units Addition ADD Subtraction SUB Multiplication MUL Division DIV Fig 9 2 Quick Mathematics menu 9 Analysis The R S HMO1002 resp R S HMO1202 series features different analysis functions f...

Страница 33: ...ve equations you can create save and retrieve a total of five different sets New equations can be added by using the universal knob to select the menu item NEW in the formula set editor Pressing the s...

Страница 34: ...e equation POWER The equation MA3 can be named ENERGY using the soft menu key LABEL Fig 9 5 Definition of the current equation Now all definitions will be completed and the results can be displayed an...

Страница 35: ...tion is useful for a precise amplitude measurement of a period signal for instance HAMMING The Hamming window function is bell shaped In contrast to the Hanning and Blackman window function it is not...

Страница 36: ...eater the selected value is the gre ater are the potential curve deviations in the amplitude The tolerance mask is displayed in white in the background The generated and edited mask can be used immedi...

Страница 37: ...e of the capacitor or inductor Fig 9 9 HZ20 adapter connected to AUX OUT If components are tested which are located in circuits or instru ments the circuits resp instruments must be disconnected first...

Страница 38: ...ement va lues of the AUTO MEASURE values The analysis function DIG VOLTMETER can be activated by using the UTIL menu in the VERTICAL control panel The Fig 9 11 Excamples of Component test digital volt...

Страница 39: ...pattern input Manual Manual switching of the 4 individual connectors UART Polarity setting the bit rate SPI I2C CAN LIN Setting the bit rate Tab 10 1 Settings options for the pattern generator 10 2 1...

Страница 40: ...dge If BOTH is selected as slope direction both edges are of equal value The first edge triggers the pattern The edges do not alternate If the MAN setting is activated the pattern is issued manually b...

Страница 41: ...emove files that are no longer required you can use the universal knob to se lect the respective settings file and remove it by pressing the soft menu key REMOVE FILE If a USB stick is connec ted you...

Страница 42: ...t menu TRACES allows you to use the universal knob in the CURSOR MENU section to select a channel which will be saved as a curve You may only select chan nels that have been activated via channel keys...

Страница 43: ...tivated waveworm is stored in color with an inverted background INVERT BG GRAY If INVERT BG GRAY is activated the waveform is stored in gray with an inverted background INVERT ALL If INVERT ALL is act...

Страница 44: ...08 logic probe to add 8 digital logic inputs All software required for the mixed signal operation is already included in the firmware of the R S HMO1002 resp R S HMO1202 series It is only necessary to...

Страница 45: ...ine the bus source and structure The contents of the menu change with the selected BUS type After pressing the top soft menu key BUS WIDTH you can use the universal knob to select a bus width from 1 8...

Страница 46: ...11 R S HV111 The option R S HOO11 resp the voucher R S HV111 can be used to trigger and decode I2C SPI and UART RS 232 buses on analog inputs only and it only allows the decoding of one serial bus at...

Страница 47: ...vate the list view blue highlighted By default the table is displayed at the bottom of the screen Generally a complete message of a protocol is displayed in a row The columns include important informa...

Страница 48: ...Slave Communication the master provides the clock pulse and selects the slave Addressing Each slave can be addressed via unique address multiple slaves can be linked with each other and can be addres...

Страница 49: ...s 112 The 10 bit addressing mode is downward compatible with the 7 bit addressing mode by using 4 of 16 reserved addresses and can be used simultaneously For a 10 bit address length a total of 1136 ad...

Страница 50: ...is necessary to set the correct logic level threshold Please refer to chapter 4 5 The default setting is 500 mV For 3 wire SPI the external trigger input will be used a CS Chip Select The threshold ca...

Страница 51: ...ecific bit sequence within the frame which start the trigger event The soft menu key BIT OFFSET Fig 13 9 Menu for the definition of a SPI bus allows you to select the first bit of the predefined bit s...

Страница 52: ...bits Even The parity bit is set to 1 if the number of ones in a specific set of bits is uneven without parity bit Odd The parity bit is set to 1 if the number of ones in a specific set of bits is even...

Страница 53: ...oft menu key SELECT BIT and the universal knob The soft menu key STATE allows you to determine the state for each bit 1 0 or X If you choose the hexadecimal input the soft menu key VALUE and the unive...

Страница 54: ...specified value If the pattern includes only 0 or 1 it is possible to trigger on an area greater than or less than the specified value The PATTERN INPUT may be binary or hexadecimal If you choose the...

Страница 55: ...c redundancy check failed NACK not Acknowledge CRC NACK cyclic redundancy check failed followed by not Acknowledge STUFF Bit Stuffing Error INS the frame is not completely contained in the acquisition...

Страница 56: ...of the identifier The correct transfer of the identifier is verified With the soft menu key SYNCHRONISATION triggering occurs if the synchronizing field indicates an error ID With the soft menu key ID...

Страница 57: ...2 USB TMC A modern alternative to the virtual COM port VCP is the control via USB TMC class TMC stands for Test Measurement Class which indicates that the connec ted measurement instrument can be reco...

Страница 58: ...ia USB in terfaces and a separate instrument class of the USB speci fication The USB TMC protocol supports service requests trigger and other GPIB specific commands The driver is included in the NI VI...

Страница 59: ...SB MTP USB MTP provides an easy solution to load data from the oscilloscope to the PC This usually does not require a driver according to Microsoft USB MTP is automatically supported for Windows XP SP...

Страница 60: ...e read only There is no way to remove the write protection memory is only useful when the acquisition has been sus pended otherwise only the screen data will be read To continue acquisition while read...

Страница 61: ...n fix statically or variable dynamically to be assigned If IP addresses in a network are assigned fix an IP address must be preset manually with each network element If IP addresses in a network are a...

Страница 62: ...HERNET as interface and press the soft menu key PA RAMETER You can specify all parameters and assign a fixed IP address You can also assign a dynamic IP address with the activated DHCP function Please...

Страница 63: ...DC AC GND Technical Data Max input voltage 1M 200Vp derates at 20db decade to 5V above 100kHz 50 R S HMO1202 series 5Veff max 30Vs Position range R S HMO1002 series R S HMO1202 series 5 Div from cente...

Страница 64: ...gnals on all analog channels R S HOO12 R S HV112 Analysis of CAN and LIN signals on analog and logic channels Trigger types by protocols I2C Start Stop ACK NACK Address Data SPI Start End Serial Patte...

Страница 65: ...ass Editing formula editor menu driven Sources all analog channels user defined constants Storage location math memory Number of formula sets 5 formula sets Number of equations 5 equations per formula...

Страница 66: ...environments Test criterion is displayed noise level within 1div for 5mV div input sensitivity Accessories included Line cord printed operating manual 2x HZ154 probes R S HMO1002 or 2x RT ZP03 probes...

Страница 67: ...1 1 Operating positions 4 Fig 1 2 Product labeling in accordance with EN 50419 6 Fig 2 1 Frontview of the R S HMO1202 8 Fig 2 2 Control panel of section A 8 Fig 2 3 Control panels of sections B C and...

Страница 68: ...ion 11 P Pass Fail test 8 36 pattern depth 39 pattern generator 39 40 peak detection 19 20 peak voltage 14 16 36 Persistence mode 27 polarity 25 37 39 54 probe attenuation detection 16 probe compensat...

Страница 69: ...69 Appendix...

Страница 70: ...70 Appendix...

Страница 71: ...71 Appendix...

Страница 72: ...w customersupport rohde schwarz com Service www service rohde schwarz com Subject to change Data without tolerance limits is not binding R S is a registered trademark of Rohde Schwarz GmbH Co KG Trade...

Отзывы: