
Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
212
User Manual 1173.9240.02 ─ 17
CALCulate<n>:PMETer<p>:RELative:STATe
<State>
This command turns relative power sensor measurements on and off.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
CALC:PMET2:REL:STAT ON
Activates the relative display of the measured value for power
sensor 2.
FETCh:PMETer<p>?
This command queries the results of power sensor measurements.
Suffix:
<p>
.
1...4
Power sensor index
Return values:
<Level>
Power level that has been measured by a power sensor.
The unit is either dBm (absolute measurements) or dB (relative
measurements).
Usage:
Query only
READ:PMETer<p>?
This command initiates a power sensor measurement and queries the results.
Suffix:
<p>
.
1...4
Power sensor index
Usage:
Query only
[SENSe:]PMETer<p>:DCYCle[:STATe]
<State>
This command turns the duty cycle correction on and off.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
PMET2:DCYC:STAT ON
Configuring the Measurement