Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
146
User Manual 1173.9286.02 ─ 15
Example:
INP:CONN:AIQI
Selects input from the analog baseband I connector.
Manual operation:
See
8.14 Automatic Measurement Configuration
The following commands are necessary to determine measurement parameters auto-
matically.
..........................................................................................146
[SENSe:]ADJust:ALL
This command initiates a measurement to determine and set the ideal settings for the
current task automatically (only once for the current measurement).
This includes:
●
Center frequency
●
Reference level
Example:
ADJ:ALL
Usage:
Event
Manual operation:
See
" Adjusting all Determinable Settings Automatically ( Auto
[SENSe:]ADJust:FREQuency
This command sets the center frequency to the frequency with the highest signal level
in the current frequency range.
Example:
ADJ:FREQ
Usage:
Event
Manual operation:
See
" Adjusting the Center Frequency Automatically ( Auto Fre-
[SENSe:]ADJust:LEVel
This command initiates a single (internal) measurement that evaluates and sets the
ideal reference level for the current input data and measurement settings. This ensures
that the settings of the RF attenuation and the reference level are optimally adjusted to
the signal level without overloading the R&S
FSW or limiting the dynamic range by an
S/N ratio that is too small.
Example:
ADJ:LEV
Automatic Measurement Configuration