Measurements and results
R&S
®
FSVA3000/ R&S
®
FSV3000
301
User Manual 1178.8520.02 ─ 08
Auto measurement
Use the
function to determine the required measurement parameters automat-
ically.
The following settings are determined:
●
Center frequency
●
RBW: determined signal bandwidth * 2
●
Time domain power limits
●
No gating
See also
Chapter 5.1, "Auto measurements"
The remote commands required to perform these tasks are described in
ter 12.5.9, "Measuring the time domain power"
Results
Activates the power results to be evaluated from the displayed trace or a limited area
of the trace.
"Peak"
Peak power over several measurements (uses trace averaging, Max
Hold)
"RMS"
RMS value from the points of the displayed trace or a segment
thereof.
"Mean"
Mean value from the points of the displayed trace or a segment
thereof. The linear mean value of the equivalent voltages is calcula-
ted.
"Std Dev"
The standard deviation of the measurement points from the mean
value.
The measurement of the mean power is automatically switched on at
the same time.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PPEak[:STATe]
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PPEak:RESult?
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:RMS[:STATe]
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:RMS:RESult?
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:MEAN[:STATe]
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:MEAN:RESult?
Limit State
Switches the limitation of the evaluation range on or off. Default setting is off.
If deactivated, the entire sweep time is evaluated. If switched on, the evaluation range
is defined by the left and right limit. If only one limit is set, it corresponds to the left limit
and the right limit is defined by the stop frequency. If the second limit is also set, it
defines the right limit.
Time domain power measurement
Содержание FSV3000
Страница 1: ...R S FSVA3000 R S FSV3000 Signal and Spectrum Analyzer User Manual 1178852002 Version 08 ÜãD2 ...
Страница 32: ...Contents R S FSVA3000 R S FSV3000 32 User Manual 1178 8520 02 08 ...
Страница 287: ...Measurements and results R S FSVA3000 R S FSV3000 287 User Manual 1178 8520 02 08 Statistical measurements APD CCDF ...
Страница 673: ...Data management R S FSVA3000 R S FSV3000 673 User Manual 1178 8520 02 08 Working with test reports ...