Measurements and results
R&S
®
FSVA3000/ R&S
®
FSV3000
256
User Manual 1178.8520.02 ─ 08
List Evaluation State (result summary)
....................................................................... 256
Saving the Result Summary (Evaluation List) to a File
............................................... 257
List Evaluation State (result summary)
Activates or deactivates the Result Summary.
Remote command:
CALCulate<n>:ESPectrum:PEAKsearch:AUTO
Show Peaks
If activated, all peaks that have been detected during an active SEM measurement are
marked with blue squares in the Spectrum diagram.
Remote command:
CALCulate<n>:ESPectrum:PEAKsearch:PSHow
Margin
Although a margin functionality is not available for the limit check, you can define a
margin (or:
threshold
) for the peak values to be displayed in the result summary. Only
peaks that exceed the margin value are displayed (also in the diagram, if activated).
Remote command:
CALCulate<n>:ESPectrum:PEAKsearch:MARGin