
Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1166
User Manual 1178.8520.02 ─ 08
BINCentered
The full measurement span is divided by the number of sweep
points to obtain
bin
s. The x-value of the sweep point is defined
as the x-value at the center of the bin (bin/2).
STARtstop
(Default): The x-value of the first sweep point corresponds to the
starting point of the full measurement span. The x-value of the
last sweep point corresponds to the end point of the full mea-
surement span. All other sweep points are divided evenly
between the first and last points.
Example:
FORM:DEXP:XDIS BINC
Manual operation:
See
FORMat:DIMPort:TRACes
<Selection>
This command selects the data to be included in a data import file (see
For details on importing data see
Chapter 7.6.3, "How to import traces"
Parameters:
<Selection>
SINGle | ALL
SINGle
Only a single trace is selected for import, namely the one speci-
fied by the
on page 1166 com-
mand.
ALL
Imports several traces at once, overwriting the existing trace
data for any active trace in the result display with the same trace
number. Data from the import file for currently not active traces
is not imported.
The <trace> parameter for the
on page 1166 command is ignored.
*RST:
SINGle
Manual operation:
See
"Import All Traces/Import to Trace"
See
MMEMory:LOAD<n>:TRACe
<Trace>, <FileName>
This command imports trace data from the specified window to an ASCII file.
Suffix:
<n>
.
Parameters:
<Trace>
Number of the trace to be stored
(This parameter is ignored for
ALL
).
Analyzing measurements (basics)
Содержание FSV3000
Страница 1: ...R S FSVA3000 R S FSV3000 Signal and Spectrum Analyzer User Manual 1178852002 Version 08 ÜãD2 ...
Страница 32: ...Contents R S FSVA3000 R S FSV3000 32 User Manual 1178 8520 02 08 ...
Страница 287: ...Measurements and results R S FSVA3000 R S FSV3000 287 User Manual 1178 8520 02 08 Statistical measurements APD CCDF ...
Страница 673: ...Data management R S FSVA3000 R S FSV3000 673 User Manual 1178 8520 02 08 Working with test reports ...