Remote Commands
R&S
®
FPL1000
555
User Manual 1178.3370.02 ─ 03
Example:
INIT:IMM;*WAI;
CALC:LIM:ACP:ACH:RES?
PASSED,PASSED
Usage:
Query only
CALCulate<n>:LIMit<k>:ACPower:ALTernate<ch>[:RELative]:STATe
This command turns the relative limit check for the alternate channels on and off.
You have to activate the general ACLR limit check before using this command with
CALCulate<n>:LIMit<k>:ACPower[:STATe]
.
Suffix:
<n>, <k>
.
irrelevant
<ch>
1 to 11
Alternate channel number
Manual operation:
See
CALCulate<n>:LIMit<k>:ACPower[:STATe]
<State>
This command turns the limit check for ACLR measurements on and off.
Suffix:
<n>, <k>
.
irrelevant
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Manual operation:
See
8.6.3.6
Performing an ACLR Measurement
The following commands are required to perform an ACLR measurement:
●
CALC:MARK:FUNC:POW:SEL ACP
●
CALCulate<n>:MARKer<m>:FUNCtion:POWer[:STATe]
●
8.6.3.7
Analyzing Measurement Results
The following commands analyze and retrieve measurement results for ACLR mea-
surements.
Useful commands for channel power measurements described elsewhere
●
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:RESult?
Configuring and Performing Measurements