Remote Control Commands
R&S
®
EX-IQ-BOX
251
User Manual 1409.5570.32 ─ 05
[:SOURce<hw>]:EBOX:SELect
<Select>
Select a currently connected R&S
EX-IQ-BOX by means of its serial number.
Parameters:
<Select>
float
Example:
SOUR:EBOX:DEV:SELect 100112
select the connected R&S
EX-IQ-BOX with serial number
10112
.
[:SOURce<hw>]:EBOX:STATe?
Queries the current state of the R&S
EX-IQ-BOX. The state can only be activated with
commands switching on a transmission protocol, e.g.
EBOX:USER:STAT ON
.
Return values:
<State>
OFF | ON
*RST:
OFF
Example:
SOUR1:EBOX:STAT?
check whether a standard is active.
Usage:
Query only
9.2.4 INSTrument Subsystem
The
INStrument
subsystem contains remote control commands to query the interface
and identity of connected R&S instruments.
The numeric suffix
INST<ch>
distinguishes between the instruments connected to the
R&S
EX-IQ-BOX. If only one R&S instrument is connected the suffix
INST[1]
is optional
and can be omitted.
.......................................................................251
........................................................................252
.......................................................................252
........................................................................252
[:SOURce<hw>]:EBOX:INST<ch>:NAME?
Queries type and serial number of an R&S instrument connected to the R&S
EX-IQ-BOX.
Return values:
<Name>
string
Example:
SOUR2:EBOX:INST1:NAME?
query the symbolic name of the connected instrument.
Response:
AMU 200A (100201)
the second connected instrument is an
R&S AMU200A
with serial
number
100201
.
Usage:
Query only