Protocol Settings
R&S
®
EX-IQ-BOX
210
User Manual 1409.5570.32 ─ 05
●
Pre-trigger
, the trigger position at the end of the waveform provides the evaluation
of the signal before the trigger event.
The trigger position is graphically displayed on the right.
Remote command:
Sample Rate
Indicates the sample rate of the signal to be recorded.
Remote command:
Recording Time
Indicates the duration of recording. The recording time results from the recording length
and the sample rate.
Remote command:
Recorder Status Diagram
Displays graphically the current process step of the recording.
"Idle"
The recording process is inactive. Configure the recorder in that state.
The recorder remains in Idle state until you start with the "Record" but-
ton.
"Free Run"
The recorder records the signal continuously in a ring memory. These
data are required for pre-trigger analyzes.
"Ready"
Recording passed through the ring memory once at least, and thus
described the memory completely. At this point the data is valid and
R&S
DigIConf activates the trigger signal automatically.
"Wait for Trigger"
The recorder continues recording until the trigger event occurs. Then
the process switches to the next state.
Note:
If trigger source "Software" is used, the trigger event occurs
immediately.
"Triggered"
After the trigger event, only the remaining samples after the trigger
position are recorded (post-trigger data).
"Record Done" If all post-trigger data are recorded, the recording is completed and the
recorder stops. All data are available in memory.
Remote command:
CPRI