Configuring the Input Signal and Measurement Mode
R&S
®
EVSG1000
51
User Manual 1178.6227.02 ─ 06
3
Configuring the Input Signal and Measure-
ment Mode
The following settings are common to all measurement modes. They define which data
is measured and how, and are indicated at the top of the measurement window.
............................................................................................... 51
................................................................................................ 52
3.1
Input Signal Settings
Access
: [Setup] > "Signal In"
Access VNC
: [n] > [F1]
The following settings configure the receiver signal to be measured by the
R&S
EVSG1000.
For some measurement modes, you can switch the type of signal input. See the
description of the individual modes.
RX2 is only available with the option R&S
EVSG-B1.
Correction RX 1 In/ Correction RX 2 In
......................................................................... 51
Correction RX 1 In/ Correction RX 2 In
Configures the correction of the RX 1/ RX 2 input signal, respectively (in dB).
("Correction RX 2 In" is only available with option R&S
EVSG-B1.)
This value is subtracted from all levels and compensates effects of upstream lines and
attenuators.
Input Signal Settings