Graphic Measurement Modes
R&S
®
EVSG1000
166
User Manual 1178.6227.02 ─ 06
Trigger mode
Determines how the trigger is evaluated.
"Norm"
Each time a trigger event occurs, a measurement is performed.
"Auto"
Each time a trigger event occurs, a measurement is performed.
If no trigger event occurs during a predefined waiting period, a mea-
surement is performed without a trigger. This helps you determine the
required trigger level.
"Single"
When the trigger event occurs, a single measurement is performed.
Trigger Source
Determines how often and when a new sweep is performed.
"Ext."
A measurement is performed each time a trigger signal from a con-
nected external trigger device is received. The trigger device must be
connected to the "Trigger In" connector on the rear panel of the
R&S
EVSG1000.
"Level"
A measurement is performed when the input signal reaches the
defined trigger level.
Trigger Edge
For external triggering, this setting defines whether a measurement is performed when
the positive (rising) or negative (falling) edge of the trigger signal is received. Select the
key repeatedly to toggle between the two settings.
Remote command:
Trigger Level
For the
= "Level", this value specifies the level in percent the input sig-
nal must reach to trigger a measurement.
Remote command:
5.4.7
Configuring the Trace
Access
: [Meas]
Access VNC
: [a]
Trace settings determine how the display values are calculated from the measured val-
ues.
AF Time Domain Mode (Option R&S
EVSG-K12)