Remote Control
R&S
®
ESRP
924
User Manual 1175.7445.02 ─ 08
[SENSe:]CORRection:TSET:SELect
<FileName>
This command creates or selects a transducer set.
Parameters:
<FileName>
String containing the name of the transducer set.
If the name does not exist yet, the R&S
ESRP creates a trans-
ducer set by that name.
Example:
See
Manual operation:
See
"Using transducer factors or transducer sets"
See
[SENSe:]CORRection:TSET[:STATe]
<State>
This command turns a transducer set on and off.
Before you can use the command, you have to select a transducer set with
.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
See
Manual operation:
See
"Using transducer factors or transducer sets"
[SENSe:]CORRection:TSET:UNIT
<Unit>
This command selects the unit of a transducer set.
Note that the unit of all transducer factors in a transducer set must be the same or in
relative terms (dB).
Before you can use the command, you have to select a transducer set with
.
Parameters:
<Unit>
String containing one of the following units:
DB | DBM | DBUV | DBUV_M | DBUA | DBUA_M | DBPW |
DBPT
*RST:
DB
Example:
See
Manual operation:
See
Using Transducers