Before you can use the command, you have to feed in a calibration signal with
DIAGnostic<n>:SERVice:INPut[:SELect]
Suffix:
<n>
.
irrelevant
Parameters:
<Frequency>
Possible frequencies of the calibration signal are fix.
If you define a frequency that is not available, the R&S
ESRP
uses the next available frequency. Example: a frequency of
20
MHz is rounded up to the next available frequency
(21.33
MHz).
*RST:
64 MHz
Example:
DIAG:SERV:INP:PULS:CFR 128 MHz
Defines a calibration signal frequency of 128 MHz.
Usage:
SCPI confirmed
Manual operation:
See
DIAGnostic<n>:SERVice:INPut[:SELect]
<Signal>
This command selects the source of the signal.
You can feed in a signal from the RF input or the calibration signal.
Suffix:
<n>
.
irrelevant
Parameters:
<Signal>
RF
RF input
CALibration
Calibration
*RST:
RF
Example:
DIAG:SERV:INP CAL
Selects the calibration signal as the input signal
Manual operation:
See
DIAGnostic<n>:SERVice:STESt:RESult?
This command queries the self test results.
Suffix:
<n>
.
irrelevant
Parameters:
<Results>
String of data containing the results.
The rows of the self test result table are separated by commas.