Remote Control
R&S
®
ESRP
899
User Manual 1175.7445.02 ─ 08
INSTrument:COUPle:SPAN
<State>
This command turns synchronization of the start and stop frequency between mea-
surement channels on and off.
Parameters:
<State>
ALL | 1
Turns on synchronization.
NONE | 0
Turns off synchronization.
*RST:
NONE
Example:
INST:COUP:SPAN ALL
Synchronizes the start and stop frequency.
11.9.4
Controlling In- and Ouputs
.................................................................................. 899
............................................................................................. 899
.................................................................................................900
...............................................................................................900
DIAGnostic<n>:SERVice:NSOurce
<State>
This command switches the 28 V supply of the noise source on the front panel on or
off.
Suffix:
<n>
.
irrelevant
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
DIAG:SERV:NSO ON
Manual operation:
See
INPut:PRESelection:STATe
<State>
This command turns the preselector on and off.
The command is available with option R&S
ESRP-B2.
Parameters:
<State>
ON | OFF
*RST:
ON