Remote Control
R&S
®
ESRP
877
User Manual 1175.7445.02 ─ 08
SOURce<n>:EXTernal<generator>:FREQuency[:FACTor]:DENominator
<Value>
This command defines the denominator of the factor with which the analyzer frequency
is multiplied in order to obtain the transmit frequency of the selected generator.
This command is only valid for the External Tracking Generator option (R&S
FSV-B10).
Select the multiplication factor such that the frequency range of the generator is not
exceeded if the following formula is applied to the start and stop frequency of the ana-
lyzer:
Offset
Analyzer
Generator
F
ator
Deno
Numerator
F
F
min
Suffix:
<n>
.
irrelevant
<generator>
1 | 2
tracking generator
For Noise Figure measurements (K30): 1 only
Parameters:
<Value>
<numeric value>
*RST:
1
Example:
SOUR:EXT:FREQ:NUM 4"
"SOUR:EXT:FREQ:DEN 3"
Sets a multiplication factor of 4/3, i.e. the transmit frequency of
the generator is 4/3 times the analyzer frequency.
Mode:
A, NF
Manual operation:
See
"Generator Frequency (Numerator/Denominator/Offset)"
SOURce<n>:EXTernal<generator>:FREQuency[:FACTor]:NUMerator
<Value>
This command defines the numerator of the factor with which the analyzer frequency is
multiplied in order to obtain the transmit frequency of the selected generator.
This command is only valid for the External Tracking Generator option (R&S
FSV-B10).
Select the multiplication factor such that the frequency range of the generator is not
exceeded if the following formula is applied to the start and stop frequency of the ana-
lyzer:
Offset
Analyzer
Generator
F
ator
Deno
Numerator
F
F
min
Suffix:
<n>
.
irrelevant
Remote Commands to Control the Tracking Generator