Remote Control
R&S
®
ESRP
679
User Manual 1175.7445.02 ─ 08
Parameters:
<Offset>
Numeric value that defines the offset.
The unit of the offset depends on the scale of the x-axis.
*RST:
0
Example:
CALC:LIM2:CONT:OFFS 100us
Sets the X offset for limit line 2 (defined in zero span) to 100µs.
Manual operation:
See
CALCulate<n>:LIMit<k>:CONTrol:SHIFt
<Distance>
This command moves a complete limit line horizontally.
Compared to defining an offset, this command actually changes the limit line definition
points by the value you define.
Suffix:
<n>
.
Selects the measurement window.
<k>
Selects the limit line.
Parameters:
<Distance>
Numeric value that defines the distance of the shift.
The unit depends on the scale of the x-axis.
Example:
CALC:LIM2:CONT:SHIF 50KHZ
Shifts all reference values of limit line 2 by 50 kHz.
CALCulate<n>:LIMit<k>:CONTrol:SPACing
<InterpolMode>
This command selects linear or logarithmic interpolation for the calculation of limit lines
from frequency points.
Suffix:
<n>
.
Selects the measurement window.
<k>
1...8
Selects the limit line.
Parameters:
<InterpolMode>
LINear | LOGarithmic
*RST:
LIN
Example:
CALC:LIM:CONT:SPAC LIN
CALCulate<n>:LIMit<k>:LOWer[:DATA]
<LimitLinePoints>
This command defines the vertical definition points of a lower limit line.
If the measured values are smaller than the LOWer limit line, the limit check is violated.
Suffix:
<n>
.
irrelevant
Remote Commands in Receiver Mode